ZX8533 RS485 ModbusRTU Command Quick Reference Tool
Applicable models: ZX8533 series; applicable firmware: V3.1.6 and above (RS485 Modbus RTU protocol supported since V3.1.6); the multi-parameter result read (register 0x70) requires the multi-parameter function form of V3.1.11 and above. This manual is used together with the "Zhixin ZX8533 Series RS485 Protocol Operation Guide". All example frames have been verified one by one through actual instrument communication and can be copied and used directly.
- Instrument Settings and Wiring
- Handshake Flow and T3.5 Frame Interval
- Frame Format (Function Code 03 / 0x10 / Exception Frame / CRC16)
- Float Big-Endian and Online Conversion Tool
- Register Map and Online Command Generator
- Reading Measurement Results (0x64 / 0x68 / 0x6E / 0x70)
- Enumeration Quick Reference (Speed/Trigger/Output Impedance/Range/Limit/Page/Status Bits)
- Notes
1. Instrument Settings and Wiring
- System Settings page → Bus Mode = RS485, Bus Address = 8 (factory default restored; first byte of the message is 0x08; range 0x01~0x20);
- Serial parameters (must match the PLC/host computer): 9600 bps (adjustable), 8 data bits, 1 stop bit, no parity;
- During RS485 communication, the instrument keypad enters a locked state (remote state); press the KEYLOCK key to unlock;
- Debugging tip: leave the page on the "System Settings" page, and the instrument will print the received command code at the bottom of the screen; if nothing is displayed at all, check whether the RS485 module was purchased.
2. Handshake Flow and T3.5 Frame Interval
2.1 A Complete Handshake
Host (PLC/host computer) Instrument (slave)
│ │
│ ① Send request frame byte by byte │
│ ────────────────────────────────────────────►│
│ [Address][Function Code][Register Addr][Count/Data][CRC] │
│ │ ② Restart T3.5 timer on each byte received
│ ③ Bus silent ≥ T3.5 (3.5 character times) │ ③ T3.5 expires → determine "one frame received"
│ ═ ═ ═ ═ ═ silent ═ ═ ═ ═ ═ │ ④ Verify: address match? CRC correct? frame complete?
│ │ no match/incorrect → discard, no response
│ │ ⑤ Execute action (write parameter/read data)
│ ⑥ Return response frame │
│ ◄────────────────────────────────────────────│
│ [Address][Function Code][Data...][CRC] │
│ or exception frame [Address][0x80+Function Code][Error Code][CRC] │
│ │
│ ⑦ Host also uses T3.5 to determine end of response frame, then sends next frame │
│ (adjacent frames must be separated by ≥ T3.5 silence) │
2.2 What Is T3.5
- T3.5 = 3.5 character times. Modbus RTU frames have no start/end markers; frame separation relies on bus silence: if silence in the middle of a byte stream exceeds T3.5, the previous frame is considered ended and the next frame begins;
- One character = 1 start bit + 8 data bits + 1 parity bit + 1 stop bit = 11 bits (this instrument counts 11 bits even with no parity);
- 9600 bps: T3.5 = 3.5 × 11 / 9600 ≈ 4.01 ms; 19200 bps ≈ 2.0 ms; when >19200, the protocol specifies a fixed 1.75 ms;
- Instrument implementation: the internal timer uses 50µs as its timing unit, reloads the count on each byte received, and determines frame end on timeout; the T3.5 timer interrupt priority is higher than the serial interrupt, ensuring frame splitting is not blocked even by long serial interrupts;
- There is also a companion T1.5: if the interval between adjacent bytes within the same frame exceeds 1.5 character times (≈ 1.72 ms at 9600), it indicates an abnormal frame interruption, and the receiver should discard the frame. Therefore, the host computer must send a complete frame continuously and without interruption; slow byte-by-byte manual sending is prohibited.
3. Frame Format (Function Code 03 / 0x10 / Exception Frame / CRC16)
3.1 Read Request (Sent by Host, Function Code 0x03)
Byte: 1 2 3 4 5 6 7 8
[Addr] [0x03] [Reg Hi ][Reg Lo ][Count Hi][Count Lo][CRC Lo ][CRC Hi ]
3.2 Read Response (Returned by Slave)
[Addr] [0x03] [Data Len M][Data 1..M ][CRC Lo][CRC Hi] Ex: 08 03 04 42 F0 00 00 77 78 → Frequency 120.0 Hz
3.3 Write Request (Sent by Host, Function Code 0x10)
Byte: 1 2 3 4 5 6 7 8..M+7 M+8 M+9
[Addr] [0x10] [Reg Hi ][Reg Lo ][Count Hi][Count Lo][Len M ][Data 1..M ][CRC Lo ][CRC Hi ]
Ex: 08 10 00 03 00 02 04 44 FA 00 00 A9 E7 → Set frequency 2000Hz
After correct execution, the slave echoes back the first 8 bytes exactly (Address + 0x10 + Register Address + Count + CRC), with no data area.
3.4 Exception Response (Returned by Slave)
[Addr] [0x80+Function Code] [Error Code] [CRC Lo] [CRC Hi]
Ex: 08 0x83 0x02 xx xx → Illegal register address/count (MB_ENOREG)
When the address does not match or the CRC is incorrect, the slave does not respond with any data.
3.5 CRC16 (Modbus Standard)
- Polynomial 0xA001 (reverse of x¹⁶+x¹⁵+x²+1), initial value 0xFFFF;
- Calculated over byte 1 through the third-to-last byte of the frame;
- Transmission order: low byte first (CRC Lo, CRC Hi).
Frame content (space-separated HEX, excluding CRC):
4. Float Big-Endian and Online Conversion Tool
Examples: 2000.0 → 44 FA 00 00; 1.2µF(1.2e-6) → 35 A1 0F B0; -1.0 → BF 80 00 00; 340.0 → 43 AA 00 00.
Float value: Unit multiplier:
HEX 4 bytes:
5. Register Map and Online Command Generator
All commands have CRCs automatically calculated according to the slave address; click "Copy" to paste into a serial assistant. To change the data area (such as the frequency value), use the tool in Section 4 to generate new data, replace it, and recalculate the CRC with the tool in Section 3.5.
5.1 Write Parameters (Function Code 0x10) — Data area is editable: numeric rows accept decimal input directly (with e exponent allowed), option rows use drop-down menus, and the command on the right recalculates CRC in real time
| Function | Register | Register Count | Data Area Content | Complete Command (HEX) | Copy |
|---|
5.2 Read Parameters/Read Results (Function Code 0x03)
| Function | Register | Register Count | Returned Data | Complete Command (HEX) | Copy |
|---|
6. Reading Measurement Results in Detail (0x64 / 0x68 / 0x6E / 0x70)
6.1 Recommended Read Cadence
Write 0x14 trigger (0001) or wait for external trigger → Poll read 0x6E status (recommended interval 50~200ms) → bit4(0x0010)=1 indicates "single-point measurement complete, data corrected" → Read 0x64 / 0x68 / 0x70 to get results → 0x6E register auto-clears after read, and is re-set in the next measurement round
6.2 Register 0x64: Primary/Secondary Parameter Results (8 Bytes)
Request: 08 03 00 64 00 04 05 4F
Response: 08 03 08 [Primary 4B][Secondary 4B] [CRC]
Example: 08 03 08 35 A0 25 69 39 84 0D 6D 0A 9E
→ Primary 35A02569 = 1.19318e-6 (1.193µF), Secondary 39840D6D = 2.5e-4
6.3 Register 0x68: Result + System Status + Bin Number (12 Bytes)
Request: 08 03 00 68 00 06 44 8D
Response: 08 03 0C [Primary 4B][Secondary 4B][sysStatus 2B][BinNo 2B] [CRC]
sysStatus: 0=Normal, 5=Contact check both H/L fail, 6=High-side fail (HNC), 7=Low-side fail (LNC)
BinNo : Sorting bin number (0=OUT, 1~n=bin number)
6.4 Register 0x6E: Operation Status Event (2 Bytes, Cleared After Read)
| Bit | Weight | Meaning |
|---|---|---|
| bit0 | 0x0001 | Correction data measurement complete (OPEN/SHORT/LOAD) |
| bit2 | 0x0004 | Curve sweep complete |
| bit3 | 0x0008 | List sweep last point complete |
| bit4 | 0x0010 | Single-point measurement complete (data corrected) — this is what "data ready" polling looks at |
| bit5 | 0x0020 | Transformer sweep complete |
08 03 02 00 10 65 89 indicates data is ready.6.5 Register 0x70: Multi-Parameter Results (26 Bytes, Customized)
Request: 08 03 00 70 00 0D 85 4D
Response: 08 03 1A [Row 1 float][Row 2 float][Row 3 float][Row 4 float][Row 5 float][Row 6 float][Overall Judgement 2B] [CRC]
Unmeasured rows = 00 00 00 00; overall judgement MP_totCmpResult: 0=PASS
7. Enumeration Quick Reference
| Parameter | Register | Value (2 Bytes) | Meaning |
|---|---|---|---|
| Speed | 0x01 | 00 00 / 00 01 / 00 02 | Fast / Medium / Slow |
| Trigger Mode | 0x02 | 00 00 / 00 01 / 00 02 / 00 04 | Internal / Manual / External / Auto |
| Measurement Function | 0x04 | 00 00 ~ 00 16 | 0=Cp-D … 22=DCR (23 items, details in table below); ZX6538 proprietary 27/28/29=Le-Q-Z / Le-Rs-Z / Ce-Rs-Le |
| Limit Mode | 0x05 | 00 00 / 00 04 | 00 00 = Percentage %; 00 04 = Absolute ABS |
| Range | 0x12 | 00 00 / 00 01 | Locked HOLD / Auto AUTO |
| Output Impedance | 0x13 | 00 00 / 00 01 / 00 02 | 30Ω / 100Ω / LOWZ |
| Trigger Command | 0x14 (write only) | 00 01 | Trigger measurement once (TrigFlag=1) |
| Bias Switch | 0x17 | 00 00 / 00 01 | Bias off / Bias on |
| Page | 0x18 | Starting from 00 00 | See page enumeration below; writing a page also saves the test conditions |
Measurement Function Enumeration (Register 0x04, 23 Items)
| Value | Function | Value | Function | Value | Function |
|---|---|---|---|---|---|
| 0 | Cp-D | 1 | Cp-Q | 2 | Cp-G |
| 3 | Cp-Rp | 4 | Cs-D | 5 | Cs-Q |
| 6 | Cs-Rs | 7 | Lp-D | 8 | Lp-Q |
| 9 | Lp-G | 10 | Lp-Rp | 11 | Ls-D |
| 12 | Ls-Q | 13 | Ls-Rs | 14 | R-X |
| 15 | Z-θdeg | 16 | Z-θrad | 17 | Rs-Q |
| 18 | G-B | 19 | Y-θdeg | 20 | Y-θrad |
| 21 | Rp-Q | 22 | DCR |
Page Enumeration Values
| Value | Page | Value | Page | Value | Page |
|---|---|---|---|---|---|
| 0 | Measurement Display LcrMeasDisp | 1 | Bin Number Display BinNoDisp | 2 | Bin Count BinCountDisp |
| 3 | List Measurement ListSweepDisp | 4 | Measurement Setup MeasSetup | 5 | User Correction Correction |
| 6 | Limit Table Setup LimitTableSetup | 7 | List Sweep Setup ListSweepSetup | 14 | System Settings SystemSetup |
8. Notes
- After writing parameters, they take effect on the hardware immediately (e.g. after frequency is written, the instrument immediately reconfigures the analog front end and refreshes the screen display);
- Writing page 0x18 triggers saving of test conditions (EEPROM); be mindful of EEPROM lifespan when switching pages frequently;
- Writing a frequency outside the MINFREQ~MAXFREQ range is ignored (no error is reported); writing a level/bias beyond the limit is clamped on the hardware side;
- 0x6E is cleared after read — with multiple masters/multi-point polling, note that only the first reader can get the event;
- Trigger mode writes accept 0/1/2/3/4/6; other values are ignored.
Version: V1.0 (2026-10-16) | CRCs of all example frames on this page are calculated in real time by the in-page tool according to the Modbus standard and can be copied and used directly