ZXP
ZX8533 Precision Calculator
Precision LCR Meter 20 Hz ~ 500 kHz
ZX8533 | Frequency range 20 Hz ~ 500 kHz | Includes DCR (DC Resistance, option) measurement accuracy calculation | Click the "Generate Heatmap" button below to plot real-time accuracy contour distribution
Measurement Function
Test Conditions
Range: 20 Hz ~ 500 kHz
Range: 0.1 mVrms ~ 2 Vrms
°C
Range: 0°C ~ 55°C
DUT Parameters
D
ZX8533 Accuracy Formula Reference
▾
Impedance Measurement Core Formula
Ae = [Ab + Zs/|Zm| x 100 + Yo x |Zm| x 100] x Kt Aa = Ae + Acal
DCR Measurement Formula (Option)
Aa = Ae + 0.03% Ae = [Ab + (Rs/|Rm| + Go x |Rm|) x 100] x Kt
SHORT Mode Ab (%)
| Frequency [Hz] | <50mV | 50mV~0.3V | 0.3V~1V | 1V~10V | 10V~20V |
|---|---|---|---|---|---|
| 20 ~ <125 | 0.60x | 0.60 | 0.30 | 0.30 | 0.30 |
| 125 ~ 1M | 0.20x | 0.20 | 0.10 | 0.15 | 0.15 |
| 1M ~ 2M | 0.40x | 0.40 | 0.20 | 0.30 | 0.30 |
MED/LONG Mode Ab (%)
| Frequency [Hz] | <30mV | 30mV~0.3V | 0.3V~1V | 1V~10V | 10V~20V |
|---|---|---|---|---|---|
| 20 ~ <100 | 0.25x | 0.25 | 0.10 | 0.15 | 0.15 |
| 100 ~ 1M | 0.10x | 0.10 | 0.05 | 0.10 | 0.15 |
| 1M ~ 2M | 0.20x | 0.20 | 0.10 | 0.20 | 0.30 |
ZX8533 Accuracy Contour Map
ZX8533 | V=1V, MED, 23°C, 0m | 400×400 | Blue = Best accuracy region | White contour lines