Short-circuit correction for automatic transformer test systems

The automatic transformer test system allows you to set three short-circuit correction frequency points—Frequency 1, Frequency 2, and Frequency 3—in the [User Correction] menu. Move the cursor to Frequency 1, Frequency 2, or Frequency 3, and use the soft keys to perform spot-frequency short-circuit correction for each of the three set frequency.

Release date:

2014-02-18

The short-circuit correction function of an automatic transformer test system is primarily designed to eliminate errors caused by parasitic impedance (R, X) in series with the device under test.

The automatic transformer test system employs the following two methods for short-circuit correction:

1. The automatic transformer test system performs short-circuit correction tests at 45 fixed frequency points, regardless of the frequency you have currently set. For frequency other than these 45 points, the short-circuit correction data are calculated using an interpolation method to determine the corresponding values across different test frequency and measurement ranges. Move the cursor to the short-circuit domain and use the soft key “Short‑Circuit Full‑Frequency Clear” to zero the short‑circuit readings over the entire frequency range. The 45 fixed frequency points follow the same procedure as described for open‑circuit correction.

2. The automatic transformer test system allows you to set three short-circuit correction frequency points—Frequency 1, Frequency 2, and Frequency 3—in the [User Correction] menu's frequency domain. Move the cursor to Frequency 1, Frequency 2, or Frequency 3, and use the soft keys to perform short-circuit correction for each of the three designated frequency.