ZX9501SA Series Hipot Tester

The ZX9501SA series hipot testers feature an 800×480 color LCD display. The output voltage magnitude, voltage rise time, test time, fall time, and output frequency are all fully controlled by the MCU. The instrument can display the breakdown current and voltage values in real time. All models in the series feature arc detection and test lead contact check functions.
  • File size: 258.6KB

  • File size: 2.0MB


Product Model

Model

Overview

Product Brochure

ZX9501SA-4 to ZX9501SA-32Dual-channel; 4/8/16/32-channel scanning function respectively; Voltage: AC 0.05–5.00 kV, DC 0.05–6.00 kV; Current: AC 0.1–30 mA, DC 0.01–10 mA; Insulation Resistance: 0.1 MΩ–10 GΩ.

Click to View

Functions & Advantages

● Chinese/English operation interface

● List display function for one-screen viewing of multi-step results

● Key lock function; automatically saves test conditions on power-off

● Open/Short Circuit Detection (OSC)

● Test lead contact check function

● Audible and visual alarm for defective products

● Fast discharge function

● Arc detection function with selectable current and level modes

● Ground Fault Interrupter (GFI) for electric shock protection

● Independently settable voltage rise time, test time, and fall time

● Supports USB flash drive and internal file storage

● Current baseline zeroing function

● Instrument software upgrade via USB flash drive

 

Stable Output Unaffected by AC Power Fluctuations

Equipped with closed-loop feedback and linear amplifier technology for precise output voltage control — stable even under load variations!

Traditional hipot testers typically use conversion transformers for direct step-up, making the output voltage susceptible to AC line input fluctuations, which may lead to inaccurate test results. When distorted voltage is applied to the DUT, it may accelerate component degradation and create hidden quality issues. The ZX9501SA series employs closed-loop output voltage control, monitoring and adjusting the output in real time to ensure voltage fluctuation within ±(1% reading + 10 V) under various DUT load conditions. The internal independent regulated power supply design effectively reduces the impact of external grid disturbances, providing a solid foundation for testing.

 

High Precision and High-Speed Measurement

The ZX9501SA series delivers high-precision measurement capability with voltage measurement accuracy of ±(1.0% reading + 5 digits) and current measurement accuracy of ±(1.0% reading + 5 digits). Equipped with auto-ranging, it achieves high-speed judgment with a minimum test time of 0.3 seconds, ensuring fast and accurate measurements.

◆  Current Zeroing Function

During testing, changes in ambient temperature and humidity as well as different cable routing can cause stray currents from distributed capacitance in the test leads and fixtures, resulting in a baseline reading under open-circuit conditions. The current zeroing function eliminates such stray current interference with one-touch, further improving measurement accuracy and ensuring reliable insulation performance evaluation.

◆  Arc Detection Function

Specifically designed to capture transient discharge pulses in the high-voltage test loop, filtering out normal current and responding only to high-speed current variations. Due to low-pass filter characteristics and the inherent randomness of arc magnitude, this function can only provide an approximate assessment of localized arcing severity.

 

Precise Time Control of Applied Voltage

Features wide-range voltage rise/fall time control! Prevents adverse effects of over-spec voltage surges on the DUT.

The ZX9501SA series supports both rise time and fall time control in AC withstand, DC withstand, and insulation resistance tests. Unlike traditional instruments that only allow short fixed rise/fall times, the ZX9501SA series allows rise or fall times to be set up to 999.9 seconds. This capability is especially useful for large capacitive loads or DUTs sensitive to voltage transients, effectively preventing insulation damage or component aging caused by transient overvoltage and ensuring tests are both rigorous and safe.

◆  Rise Time Control Function

Users can set the time for voltage to rise from 0 to the set value based on DUT characteristics, adjustable from 0.1 s to 999.9 s (can be disabled).

◆  Fall Time Control Function

After the test, users can set the time for voltage to smoothly decrease from the set value to 0, adjustable from 0.1 s to 999.9 s (can be disabled).

 

Active Safety Protection Mechanisms

◆  Ground Fault Interrupter (GFI)

Monitors ground return current in real time. If the threshold of 0.5 mA is exceeded, the instrument forcibly cuts off high-voltage output and exits the test state. Even if the operator disables this function, the instrument will still automatically trigger protection when the ground current exceeds 70 mA, ensuring no serious injury to the operator. Provides active, reliable safety protection for operators.

◆  Interlock Switch (INTERLOCK)

INTERLOCK is used for multi-unit interconnection, preventing damage caused by simultaneous output from multiple instruments due to misoperation. When the interlock switch is cut off, the instrument forcibly inhibits high-voltage startup. In other words, the interlock loop being closed is a prerequisite for the instrument to output high voltage.

◆  Forced Fast Discharge Function

After DC withstand or insulation resistance testing, the instrument automatically discharges high-voltage charge stored in capacitive loads to a safe voltage level, preventing residual electric shock hazards.

◆  Open/Short Circuit Detection (OSC)

Pre-tests the test loop before high-voltage application, capable of detecting distributed impedance currents above 100 pF to confirm reliable connections, preventing misjudgment caused by open test leads or poor pin contact, and ensuring accurate and safe high-voltage testing.

 

Multi-Step Results on One Screen

◆  List Sweep Function

Users can freely edit multiple measurement steps based on test requirements, with each step independently configurable for test voltage, test time, upper/lower limits, and other parameters. Enables one-screen display of measurement results from different steps and conditions.

 

PC Communication Software

The companion software communicates with the instrument via RS-232C, supporting parameter setting, test initiation, and data monitoring. Test results can be exported to Excel format for quality traceability and data analysis.

 

Wide Range of Applications

◆  Power Equipment and Grid Maintenance

◆  Cable and Transformer Manufacturing

◆  New Energy Sector

◆  Industrial Electrical Equipment

◆  Rail Transit and Aviation

◆  Laboratories and Certification Bodies

Parameters & Dimensions

ModelZX9501ZX9501SZX9501BZX9501CZX9502
Withstanding Voltage Test
Output VoltageAC

0.05 kV to 5 kV ±(1.0% of reading + 5 digits) 

(50 Hz and 60 Hz selectable)

0.05 kV to 10 kV ±(1.0% of reading + 5 digits) 

(50 Hz and 60 Hz selectable)

DC0.05 kV to 6 kV ±(1.0% of reading + 5 digits)------0.05 kV to 10 kV ±(1.0% of reading + 5 digits)
Voltage Regulation Rate≤(1.0% + 10 V)(Rated Power)
Current Test RangeAC0.1 mA to 30 mA0.1 mA to 20 mA0.1 mA to 20 mA
DC0.01 mA to 10 mA0.01 mA to 5 mA------0.1 mA to 5 mA
Test Accuracy±(1.0% of reading + 5 digits)------±(1.0% of reading + 5 digits)
Discharge FunctionAutomatic discharge after test (DCW)------Automatic discharge after test (DCW)
Insulation Resistance Test
Output Voltage0.05 kV to 1 kV ±(1.0% of reading + 2 V)------0.05 kV to 1 kV ±(1.0% of reading + 2 V)
Resistance Measurement Range0.1 MΩ to 10 GΩ (Current range: 10 nA to 10 mA)------0.1 MΩ to 10 GΩ (Current range: 10 nA to 10 mA)
Resistance Measurement Accuracy500 V to 1000 V1 MΩ to 1 GΩ ±(5% of reading + 5 digits)------1 MΩ to 1 GΩ ±(5% of reading + 5 digits)
1 GΩ to 10 GΩ ±(10% of reading + 5 digits)1 GΩ to 10 GΩ ±(10% of reading + 5 digits)
50 V to 500 V0.1 MΩ to 1 GΩ ±(10% of reading + 5 digits)------0.1 MΩ to 1 GΩ ±(10% of reading + 5 digits)
Discharge FunctionAutomatic discharge after test------Automatic discharge after test
Arc Detection
Measurement RangeACLevels 1–9 (Factory default: 5)
(Corresponding current: 20 mA, 18 mA, 16 mA, 14 mA, 12 mA, 10 mA, 7.7 mA, 5.5 mA, 2.8 mA)
DCLevels 1–9Levels 1–9------Levels 1–9
General Parameters
8-Channel Matrix Scanner------Yes------------
Memory120 sets, 100 steps per set
Voltage Rise Time0.1 s to 999 s
Voltage Fall Time0.1 s to 999 s
Voltage Dwell Time0.1 s to 99.9 s (DC only)
Test Time Setting0.3 s to 999 s
Interface
StandardRS-232C, USB, HANDLER, REMOTE I/O
OptionalGPIB, SCAN

Unit: millimeters (mm)

Accessories & Options

ZX95Y06 Test Cable

ZX95Y06 Test Cable

ZX95Y-24CH/32CH/64CH

ZX95Y-24CH/32CH/64CH

Software & Resources

Changzhou Zhixin ZX95 & ZX93 Series Hipot Tester SCPI Communication Command Set V1.7

2026-08-12

ZX93 & ZX95 Series PC Software V0.3.0

2026-08-21

ZX9501_ZX9320 RS485 MODBUS Protocol Operation Guide V1.3

2026-11-12

Solutions

undefined

undefined

Customer Q&A

Q: What should be done if arcing or sparking occurs during measurement?

A: This can be mitigated by adding ferrite cores to the test leads and ensuring reliable grounding of the instrument chassis.

Q: Why is there a current reading or an abnormally low resistance value during an open-circuit test after the fixture is connected?

A: Ensure that the test fixture itself has good insulation. If the reading returns to normal after removing all test leads and performing an open-circuit test, the issue is most likely due to insufficient insulation of the test fixture. For minor current offset issues, the baseline current can be cleared using the zeroing function.

Q: For the version with Contact Check (CK) function, what could be the reasons if the CK test consistently fails?

A: ▪ CK check is only applicable when both the high-side and low-side contact surfaces are conductive. If either side is insulated, the contact check function cannot be used.

   ▪ The two contact probes of the channel may not be making reliable contact with the conductive surface of the DUT.

Total 3 < 1 >

Enter information to retrieve data.

By systematically retrieving and analyzing relevant input information, comprehensive and specialized data are obtained to support or meet specific needs or objectives.

Confirm
%{tishi_zhanwei}%