ZX2736 Series Magnetic Core Characteristic Tester
Category:
-
File size: 292.2KB
-
File size: 5.5MB
-
File size: 999.7KB
Product Model
Model | Overview | Product Brochure |
| ZX2736 | Test Current: 1 mA–300 mA; Measurement Range: 1 μV–500 mV; Supports sine wave. | Click to View |
| ZX2736-05 | Test Current: 1 mA–500 mA; Measurement Range: 1 μV–800 mV; Supports sine wave, half-wave, and full-wave. | Click to View |
| ZX2736-10 | Test Current: 1 mA–1 A; Measurement Range: 1 μV–1.0 V; Supports sine wave, half-wave, and full-wave. | Click to View |
Functions & Advantages
● Three selectable modes: Average, True RMS, and ABX-7, enabling accurate measurement of the true RMS value of sine waves and distorted waveforms such as half-wave and full-wave ● Can be used as a low-voltage AC constant current source or an AC millivoltmeter ● Fast measurement speed, high accuracy, and excellent stability ● PASS/FAIL dual-color LED indicators with multi-state buzzer alerts ● Selectable speed and ranges to meet diverse testing requirements | ● Simplified Chinese interface, easy to learn and operate ● Key lock function to prevent misoperation ● Automatically saves current test conditions on power-off, no reconfiguration needed on next startup ● Test conditions can be saved as 50 sets internally and 500 sets on USB flash drive ● 14-bin PASS sorting, with HI and LOW limit judgment |
Test Principle
The instrument operates based on the volt-ampere (V-A) method. During testing, a precisely adjustable AC constant current source is applied through the Hcur/Lcur terminals to a coil wound around the DUT core, exciting the core to generate an alternating magnetic field; meanwhile, the induced AC voltage on the core is measured through the Hpot/Lpot terminals. According to Ohm's law Vx = I × Z , with the excitation current I known, the measured voltage Vx directly reflects the equivalent impedance Z of the core at that operating point. Therefore, by comparing whether the induced voltage V falls within the preset range, the volt-ampere characteristics of the core can be judged as acceptable or not.

Excitation Output Function
◆ AC Constant Current Source
Output current is continuously adjustable from 1 mA: ZX2736 is 1 mA–300 mA, ZX2736-05 is 1 mA–500 mA, and ZX2736-10 is 1 mA–1 A. The three current specifications adapt to the testing requirements of different cores, maintaining 1% current accuracy across the full range to ensure consistency of test conditions.

◆ Excitation Waveform
The ZX2736 supports sine wave; the ZX2736-05/-10 supports sine wave, half-wave, and full-wave. Sine wave is suitable for conventional core volt-ampere characteristic testing, providing a standardized evaluation reference; half-wave and full-wave excitation can be used to evaluate the response characteristics of cores under non-sinusoidal excitation conditions. The harmonic components of the induced signal differ from those under sine wave excitation, providing a more comprehensive reflection of core performance differences under various excitation conditions and meeting the inspection requirements of specific application scenarios such as residual current protection transformers.

◆ Test Frequency
Two test frequencies, 50 Hz and 60 Hz, are selectable, providing standardized excitation for core characteristic evaluation under power frequency conditions and ensuring reliable reference benchmarks for consistency control.
◆ Half-Wave/Full-Wave Current Output Modes
For half-wave and full-wave modes, two current output modes are provided: "Before Rectification" and "After Rectification".

Before Rectification: The instrument outputs the original waveform current with the negative half-cycle removed (half-wave) or inverted (full-wave); this current value is not equal to the set value.
After Rectification: The instrument optimizes and controls the output waveform so that the effective current applied to the DUT is exactly equal to the set value.
Voltage Measurement Function
◆ AC Millivoltmeter
Minimum resolution of 1 μV; the ZX2736 has a measurement range of 1 μV–500 mV, the ZX2736-05 extends to 1 μV–800 mV, and the ZX2736-10 extends to 1 μV–1.0 V, with voltage accuracy of 1% ± 3 μV. The wide coverage from μV-level weak signals to conventional ranges balances precise small-signal measurement with conventional core testing requirements.

◆ Measurement Modes
To evaluate instrument performance, it is recommended to use high-end benchtop multimeters from Keysight Technologies or Keithley. Handheld multimeters commonly available on the market generally cannot meet the requirements.
Three modes are selectable: Average (AVG), True RMS, and ABX-7.

Average (AVG): Treats the input waveform as a quasi-sine wave and performs rectified averaging on the AC signal before calculation. This mode is suitable for fast measurement of sinusoidal signals; if the waveform is distorted, the error will be larger. This mode can be selected when comparing with low-end handheld multimeters.
True RMS: Performs root-mean-square calculation on the induced waveform according to the effective value formula, directly calculating and displaying the true effective value of waveforms containing harmonic components. The measurement result is independent of waveform shape. This mode can be selected when comparing with high-end multimeters such as the Keysight 34460A and Keithley 2001.
ABX-7: Uses a non-rectified average value algorithm for compatibility with early instruments. It is not a general-purpose measurement mode and is suitable for specific production line testing standards.
◆ Four Ranges
Select the range closest to the measured signal amplitude to ensure the measurement result falls within the optimal accuracy range.

Sorting Judgment and Result Output
◆ Comparator Function
Provides 14-bin PASS sorting (BIN1–BIN14) along with HI (above upper limit) and LOW (below lower limit) judgment. It can compare measurement results with preset standard values or ranges, display and output judgment results, and also output via EXT I/O.

◆ Deviation Modes
Three deviation modes are provided, allowing users to flexibly choose according to production line judgment requirements.

◆ Multi-State Buzzer Alert Function
Test results have two audible alerts: PASS and FAIL. The sounds differ clearly in frequency, pitch, or rhythm, ensuring quick identification.
◆ Comparator Indicator Lights
Judgment results are displayed via front-panel PASS (green) and FAIL (red) dual-color LED indicators. You can work without looking at the screen, improving work efficiency.
Parameters & Dimensions
| Model | ZX2736 | ZX2736-05 | ZX2736-10 |
| Test Frequency | 50 Hz, 60 Hz | ||
| Measurement Function | Sine wave | Sine wave, Half wave, Full wave | |
| Measurement Range | 1 μV to 500 mV | 1 μV to 800 mV | 1 μV to 1.0 V |
| Display Range | 0.001 mV to 499.99 mV | 0.001 mV to 799.99 mV | 0.001 mV to 999.99 mV |
| Test Speed | Fast: Approx. 40 meas/sec; Medium: Approx. 20 meas/sec; Slow: Approx. 6 meas/sec | ||
| Test Current | 1 mA to 300 mA, accuracy 1% | 1 mA to 500 mA, accuracy 1% | 1 mA to 1 A, accuracy 1% |
| Voltage Accuracy | 1% (±3 μV) | ||
| Range Mode | 500 mV, 200 mV, 20 mV, 3 mV | 800 mV, 200 mV, 20 mV, 3 mV | 1 V, 200 mV, 20 mV, 3 mV |
| Measurement Mode | Average, True RMS, ABX-7 | ||
| Wide Application Range | Can be used as an AC constant current source and AC millivoltmeter | ||
| Deviation Mode | Direct read, absolute deviation, coefficient | ||
| Comparator | BIN1–BIN14, HI, LOW | ||
| PASS, FAIL (LED indicator), Buzzer Alert | |||
| Trigger Mode | Internal, Manual, External | ||
| Interface | GPIB (optional), RS-232C, USB Host, USB Device, Handler | ||
| Memory | Internal: 50 sets External USB flash drive: 500 sets | ||
| Temperature & Humidity | 0 °C to 40 °C, Relative Humidity ≤ 75% | ||
| Power Requirements | 100 to 120 VAC or 198 to 242 VAC | ||
| 46 to 64 Hz, Power >50 VA | |||
| Dimensions (W × H × D) | 216 mm × 87 mm × 300 mm | ||
| Weight | 3.2 kg | ||

Unit: millimeters (mm)
Accessories & Options
ZX2735Y01 Test Platform
ZX2736Y02 Test Cable
Software & Resources
Solutions