Product Categories
ZX8538 Precision LCR Meter
Category:
-
File size: 7.0MB
-
File size: 545.6KB
-
File size: 5.0MB
Product Model
Model | Overview | Product Brochure |
ZX8538 | Basic Accuracy: ±0.05%; Frequency: 20 Hz–1 MHz; Test Level: 5 mV–2 V; Four-terminal pair auto-balancing bridge technology; Multi-point list sweep; Multi-parameter graphical analysis. | |
ZX8538A | Basic Accuracy: ±0.05%; Frequency: 20 Hz–2 MHz; Test Level: 5 mV–2 V; Four-terminal pair auto-balancing bridge technology; Multi-point list sweep; Multi-parameter graphical analysis. | Click to View |
ZX8538H | Basic Accuracy: ±0.05%; Frequency: 20 Hz–2 MHz; Test Level: 5 mV–20 V; DC Bias: 0–40 V; Four-terminal pair auto-balancing bridge technology; Multi-point list sweep; Multi-parameter graphical analysis. | Click to View |
Functions & Advantages
● Auto-balancing bridge technology, four-terminal pair Kelvin test terminals ● Provides internal DC bias: -40 V to +40 V ● Simplified Chinese and English operation languages ● Graphical sweep, open/short/load correction, and other functions | ● Save test data and test conditions (via USB flash drive or internal memory) ● Direct screen capture of graphical sweep images to USB flash drive ● Enhanced test terminal protection function ● Host computer interfaces: USB, GPIB (optional), RS-232C, LAN (optional) |
Auto-Balancing Bridge Test Principle
The ZX8538 employs the auto-balancing bridge test principle: the traditional simple operational amplifier measurement architecture has been comprehensively upgraded to a vector negative feedback network composed of core modules including the Lpot null detection circuit, 0°/90° digital phase detection and modulation/demodulation circuits, and the Lcur high-frequency drive circuit. This architectural innovation brings the "virtual ground" terminal of the measurement circuit closer to the ideal state, significantly enhancing the instrument's measurement accuracy and anti-interference capability across various practical application scenarios, and meeting the measurement requirements for most low-voltage parameters of components and materials.

Expanded Measurable Material Characteristics
◆ Fast and Accurate Dielectric Constant Measurement for Solid Materials
In recent years, the rapid development of electronic and power technologies has made the material characteristics of electronic components a key factor affecting circuit behavior. In the manufacturing of high-capacitance Multilayer Ceramic Capacitors (MLCCs), which are increasingly widely used, high dielectric constant materials are required. Before material selection, comprehensive electrical performance evaluations must be conducted. The ZX8538 has optimized its fundamental performance and can be paired with our dedicated dielectric constant test fixture ZX85Y77(L) to enable fast and accurate measurement of the capacitance (C) of solid materials, automatically calculate the material's dielectric constant (κ) via formula, and stably display the material's dissipation factor (D).

In core application fields such as integrated circuits, power devices, optoelectronic devices, and memory chips, semiconductor C-V characteristics are fundamental to reliability from the material and process levels to the device level, and are also crucial characteristics of MOS transistors and power devices. Therefore, guided by the need to measure this performance, the ZX8538 is specifically designed with a graphical sweep function, enabling users to perform convenient and intuitive data analysis while rapidly and accurately measuring these characteristics.

Typical Semiconductor Measurement Application Scenarios
◆ Diode Measurement


◆ MOSFET Measurement


Intuitive Display on Large LCD Screen
◆ Graphical Sweep Page
The instrument's large LCD screen can directly display graphical sweep curves without the need for an external computer. Typical values can be quickly and accurately located using the cursor keys, facilitating intuitive data analysis for users.

Extended Instrument Service Life
◆ USB Data Storage Function
Measurement data can be conveniently saved to a USB flash drive and transferred to a computer for result analysis, reducing space occupation on the production line.

Parameters & Dimensions
Model | ZX8538 | ZX8538A | ZX8538H | |
Test Level | 0.005-2 Vrms | 0.005-20 Vrms(<=1 MHz), 0.005-15 Vrms(>1 MHz) | ||
Test Frequency | 20 Hz to 1 MHz | 20 Hz to 2 MHz | 20 Hz to 2 MHz | |
Test Parameters | C, L, X, B, R, G, D, Q, ESR, Rp, θ, |Z|, |Y| | |||
Basic Accuracy | 0.05%; refer to the accuracy table | |||
Test Speed | Fast / Medium / Slow ; Refer to the table below for specific measurement times | |||
| Display Range | |Z|, R, X | 0.01 mΩ to 99.9999 MΩ | ||
|Y|, G, B | 0.00001 μS to 99.9999 S | |||
C | 0.00001 pF to 9.99999 F | |||
D | 0.00001 to 9.99999 | |||
L | 0.00001 μH to 99999.9 H | |||
Q | 0.01 to 99999.9 | |||
θ(RAD) | -3.14159 to 3.14159 | |||
θ(DEG) | -179.999° to 179.999° | |||
Δ% | -999.999% to 999.999% | |||
Internal DC Bias Voltage | -5 V to +5 V | -5 V to +5 V | -40 V to +40 V | |
Internal Bias Option | Optional configuration: ±100 mA, -40 V to +40 V | |||
Range Mode | Auto, Hold | |||
Trigger Mode | Internal, Manual, Auto, External, Bus | |||
Correction Function | Open/Short, All-frequency zeroing, Spot-frequency zeroing | |||
Equivalent Circuit | Series, Parallel | |||
Comparator | 10-bin sorting, BIN0–BIN9, NG, AUX | |||
PASS, HI, LOW (PASS/FAIL LED indication) | ||||
Interface | RS-232C, USB Host, USB Device, Handler, GPIB (optional) | |||
Temperature & Humidity | 0 °C to 40 °C, Relative Humidity ≤ 75% | |||
Power Requirements | 100 to 120 VAC or 198 to 242 VAC | |||
46 to 64 Hz, power > 75 VA | ||||
Dimensions (W × H × D) | 400 mm × 132 mm × 410 mm | |||
Weight (net weight) | 8.5 kg | |||
◆ Appendix: Measurement Time [ms] (DC Bias: OFF)
| Speed Mode | Test Frequency | ||||||
| 20 Hz | 100 Hz | 1 kHz | 10 kHz | 100 kHz | 1 MHz | 2 MHz | |
| Slow | 480 | 300 | 240 | 230 | 220 | 220 | 220 |
| Medium | 380 | 180 | 110 | 92 | 89 | 88 | 88 |
| Fast | 330 | 100 | 20 | 7.7 | 5.7 | 5.6 | 5.6 |

Unit: millimeters (mm)
Accessories & Options
ZX28Y05A Kelvin Test Fixture
ZX28Y10
ZX85Y47 Kelvin Test Fixture
ZX28Y11AS Four-Terminal Pair (4TP) Test Cable
ZX28Y09 SMD Kelvin Test Cable
ZX28Y08 SMD Test Fixture
Software & Resources
Customer Q&A
Q: When measuring on the single measurement page, the secondary parameter on the second line is clearly outside the upper/lower limits, yet the instrument still displays PASS. Why?
A: If the secondary parameter needs to be included in the judgment, you must enable the "Auxiliary" option in the binning settings.
Q: In the graphical sweep function, the resonant frequency and impedance readings are incorrect. What should I do?
A: The graphical sweep typically requires an all-frequency zeroing. When using test leads, ensure that the lead position during short zeroing is exactly the same as during measurement, to avoid deviations caused by lead movement.
Q: In the graphical sweep function, only one curve is displayed, making it difficult to observe. How can this be resolved?
A: In the graphical sweep settings, change the display mode to "A+B" and set the Y-axis scale to "LOG".
Q: Why does the instrument not alarm on defective products, or why does it keep beeping constantly?
A: ▪ Make sure the upper/lower limits and nominal value are configured, and that the comparator is turned ON.
▪ Check the beep settings under "System". Set "Pass Beep" and "Fail Beep" to the desired mode.
Q: Why do the readings fluctuate significantly and become unstable during zeroing?
A: When shorting the clips, ensure that the corresponding color-matched clips (red-to-blue and black-to-white) are clipped on the same side.
Q: When connecting to a PLC, the instrument output signal is found to be invalid. How can this be resolved?
A: The Handler interface output mode is set to "Hold" by default. You can change it to "Delay Clear" or "Start Clear" mode.
Q: Why do the measured values differ significantly from the standard values?
A: ▪ Check whether the short-circuit reading (pay attention to the clip connection method) is near 0 mΩ. Re-zero the instrument whenever the test environment changes, such as after replacing the fixture or test leads.
▪ Check whether compensation has been applied. Try disabling compensation and re-zeroing.
▪ If the test leads are long, have multiple adapters, or are aged, inspect all four test leads for open circuits or poor contact.
▪ Verify that the instrument's range matches the DUT. A large range mismatch can cause measurement errors.