ZX8538 Precision LCR Meter

The ZX8538A Precision LCR Meter utilizes auto-balancing bridge technology and four-terminal pair Kelvin test terminals, with a measurement frequency range covering 20 Hz to 2 MHz, a minimum resolution of 1 mHz, and a basic measurement accuracy of up to 0.05%. Its high precision and wide frequency range can meet the measurement requirements for most low-voltage parameters of components and materials. It is widely applicable for impedance performance research and analysis, as well as production line QC inspection of components such as inductors, capacitors, resistors, microphones, resonators, liquid crystal displays, varactor diodes, and transformers. The ZX8538H can be used to test the barrier capacitance (C) of PN junctions in semiconductor devices under varying bias voltages (V). In addition to connecting to a PC host software for scanning and analysis, it also provides the capability to directly display C-V sweep curve analysis on the instrument's large LCD screen. This eliminates the need for customers to configure a separate PC for each instrument, ensuring testing efficiency while reducing costs and facilitating use on production lines. The graphical displays and test data shown on the screen can also be directly copied to a USB flash drive. This instrument can also test parameters such as the input capacitance (Ciss) and gate resistance (Rg) of MOS transistors. Additionally, it features a built-in ±10 VDC voltage source to facilitate the testing of three-terminal devices.
  • File size: 7.0MB

  • File size: 545.6KB

  • File size: 5.0MB


Product Model

Model

Overview

Product Brochure

ZX8538

Basic Accuracy: ±0.05%; Frequency: 20 Hz–1 MHz; Test Level: 5 mV–2 V; Four-terminal pair auto-balancing bridge technology; Multi-point list sweep; Multi-parameter graphical analysis.

Click to View

ZX8538A

Basic Accuracy: ±0.05%; Frequency: 20 Hz–2 MHz; Test Level: 5 mV–2 V; Four-terminal pair auto-balancing bridge technology; Multi-point list sweep; Multi-parameter graphical analysis.Click to View

ZX8538H

Basic Accuracy: ±0.05%; Frequency: 20 Hz–2 MHz; Test Level: 5 mV–20 V; DC Bias: 0–40 V; Four-terminal pair auto-balancing bridge technology; Multi-point list sweep; Multi-parameter graphical analysis.Click to View

Functions & Advantages

● Auto-balancing bridge technology, four-terminal pair Kelvin test terminals  

● Provides internal DC bias: -40 V to +40 V  

● Simplified Chinese and English operation languages  

● Graphical sweep, open/short/load correction, and other functions

● Save test data and test conditions (via USB flash drive or internal memory)  

● Direct screen capture of graphical sweep images to USB flash drive  

● Enhanced test terminal protection function  

● Host computer interfaces: USB, GPIB (optional), RS-232C, LAN (optional)

 

Auto-Balancing Bridge Test Principle

The ZX8538 employs the auto-balancing bridge test principle: the traditional simple operational amplifier measurement architecture has been comprehensively upgraded to a vector negative feedback network composed of core modules including the Lpot null detection circuit, 0°/90° digital phase detection and modulation/demodulation circuits, and the Lcur high-frequency drive circuit. This architectural innovation brings the "virtual ground" terminal of the measurement circuit closer to the ideal state, significantly enhancing the instrument's measurement accuracy and anti-interference capability across various practical application scenarios, and meeting the measurement requirements for most low-voltage parameters of components and materials.

 

Expanded Measurable Material Characteristics

◆  Fast and Accurate Dielectric Constant Measurement for Solid Materials

In recent years, the rapid development of electronic and power technologies has made the material characteristics of electronic components a key factor affecting circuit behavior. In the manufacturing of high-capacitance Multilayer Ceramic Capacitors (MLCCs), which are increasingly widely used, high dielectric constant materials are required. Before material selection, comprehensive electrical performance evaluations must be conducted. The ZX8538 has optimized its fundamental performance and can be paired with our dedicated dielectric constant test fixture ZX85Y77(L) to enable fast and accurate measurement of the capacitance (C) of solid materials, automatically calculate the material's dielectric constant (κ) via formula, and stably display the material's dissipation factor (D).

In core application fields such as integrated circuits, power devices, optoelectronic devices, and memory chips, semiconductor C-V characteristics are fundamental to reliability from the material and process levels to the device level, and are also crucial characteristics of MOS transistors and power devices. Therefore, guided by the need to measure this performance, the ZX8538 is specifically designed with a graphical sweep function, enabling users to perform convenient and intuitive data analysis while rapidly and accurately measuring these characteristics.

 

Typical Semiconductor Measurement Application Scenarios

◆  Diode Measurement

 

◆  MOSFET Measurement

 

 

Intuitive Display on Large LCD Screen

◆  Graphical Sweep Page

The instrument's large LCD screen can directly display graphical sweep curves without the need for an external computer. Typical values can be quickly and accurately located using the cursor keys, facilitating intuitive data analysis for users.

 

Extended Instrument Service Life

◆  USB Data Storage Function

Measurement data can be conveniently saved to a USB flash drive and transferred to a computer for result analysis, reducing space occupation on the production line.

Parameters & Dimensions

Model

ZX8538

ZX8538A

ZX8538H

Test Level

0.005-2 Vrms

0.005-20 Vrms(<=1 MHz), 0.005-15 Vrms(>1 MHz)

Test Frequency

20 Hz to 1 MHz

20 Hz to 2 MHz

20 Hz to 2 MHz

Test Parameters

C, L, X, B, R, G, D, Q, ESR, Rp, θ, |Z|, |Y|

Basic Accuracy

0.05%; refer to the accuracy table

Test Speed

Fast / Medium / Slow ; Refer to the table below for specific measurement times

Display Range

|Z|, R, X

0.01 mΩ to 99.9999 MΩ

|Y|, G, B

0.00001 μS to 99.9999 S

C

0.00001 pF to 9.99999 F

D

0.00001 to 9.99999

L

0.00001 μH to 99999.9 H

Q

0.01 to 99999.9

θ(RAD)

-3.14159 to 3.14159

θ(DEG)

-179.999° to 179.999°

Δ%

-999.999% to 999.999%

Internal DC Bias Voltage

-5 V to +5 V

-5 V to +5 V

-40 V to +40 V

Internal Bias Option

Optional configuration: ±100 mA, -40 V to +40 V

Range Mode

Auto, Hold

Trigger Mode

Internal, Manual, Auto, External, Bus

Correction Function

Open/Short, All-frequency zeroing, Spot-frequency zeroing

Equivalent Circuit

Series, Parallel

Comparator

10-bin sorting, BIN0–BIN9, NG, AUX

PASS, HI, LOW (PASS/FAIL LED indication)

Interface

RS-232C, USB Host, USB Device, Handler, GPIB (optional)

Temperature & Humidity

0 °C to 40 °C, Relative Humidity ≤ 75%

Power Requirements

100 to 120 VAC or 198 to 242 VAC

46 to 64 Hz, power > 75 VA

Dimensions (W × H × D)

400 mm × 132 mm × 410 mm

Weight (net weight)

8.5 kg

◆  Appendix: Measurement Time [ms] (DC Bias: OFF)

Speed ModeTest Frequency
20 Hz100 Hz1 kHz10 kHz100 kHz1 MHz2 MHz
Slow480300240230220220220
Medium38018011092898888
Fast330100207.75.75.65.6

 

Unit: millimeters (mm)

Accessories & Options

ZX28Y05A Kelvin Test Fixture

ZX28Y05A Kelvin Test Fixture

ZX28Y10

ZX28Y10

ZX85Y47 Kelvin Test Fixture

ZX85Y47 Kelvin Test Fixture

ZX28Y11AS Four-Terminal Pair (4TP) Test Cable

ZX28Y11AS Four-Terminal Pair (4TP) Test Cable

ZX28Y09 SMD Kelvin Test Cable

ZX28Y09 SMD Kelvin Test Cable

ZX28Y08 SMD Test Fixture

ZX28Y08 SMD Test Fixture

Software & Resources

Zhixin Precision LCR PC General Demo Software Installer

2026-07-20

Customer Q&A

Q: When measuring on the single measurement page, the secondary parameter on the second line is clearly outside the upper/lower limits, yet the instrument still displays PASS. Why?

A: If the secondary parameter needs to be included in the judgment, you must enable the "Auxiliary" option in the binning settings.

Q: In the graphical sweep function, the resonant frequency and impedance readings are incorrect. What should I do?

A: The graphical sweep typically requires an all-frequency zeroing. When using test leads, ensure that the lead position during short zeroing is exactly the same as during measurement, to avoid deviations caused by lead movement.

Q: In the graphical sweep function, only one curve is displayed, making it difficult to observe. How can this be resolved?

A: In the graphical sweep settings, change the display mode to "A+B" and set the Y-axis scale to "LOG".

Q: Why does the instrument not alarm on defective products, or why does it keep beeping constantly?

A: ▪ Make sure the upper/lower limits and nominal value are configured, and that the comparator is turned ON.
   ▪ Check the beep settings under "System". Set "Pass Beep" and "Fail Beep" to the desired mode.

Q: Why do the readings fluctuate significantly and become unstable during zeroing?

A: When shorting the clips, ensure that the corresponding color-matched clips (red-to-blue and black-to-white) are clipped on the same side.

Q: When connecting to a PLC, the instrument output signal is found to be invalid. How can this be resolved?

A: The Handler interface output mode is set to "Hold" by default. You can change it to "Delay Clear" or "Start Clear" mode.

Q: Why do the measured values differ significantly from the standard values?

A: ▪ Check whether the short-circuit reading (pay attention to the clip connection method) is near 0 mΩ. Re-zero the instrument whenever the test environment changes, such as after replacing the fixture or test leads.

   ▪ Check whether compensation has been applied. Try disabling compensation and re-zeroing.

   ▪ If the test leads are long, have multiple adapters, or are aged, inspect all four test leads for open circuits or poor contact.

   ▪ Verify that the instrument's range matches the DUT. A large range mismatch can cause measurement errors.