ZX8816 Low Inductance Impulse Winding Tester

The ZX8816 features a 1000 Vdc pulse voltage and a 200 MHz high-speed sampling rate, capable of testing inductance products in the range of 0.1 µH to 1 mH. It offers multiple judgment functions including Area Size Comparison, Differential Area Comparison, Flutter Value Comparison, Laplacian Value Comparison, Peak Ratio Comparison, Δ Peak Ratio Comparison, and Δ Resonant Area Comparison, enabling effective detection of insulation defects in coils and meeting the testing requirements for the majority of power inductors.
  • File size: 5.6MB

  • File size: 353.8KB

  • File size: 2.3MB


Product Model

Model

Overview

Product Brochure

ZX8816

Pulse Voltage Output: 10–1000 V; Waveform Sampling Rate: 200 Msps; Minimum Measurable Inductance: 0.1 µH; Pass/Fail Audible and Visual Alarm Output; Interfaces: Handler, RS-232C, USB Host.

Click to View

Functions & Advantages

● Pulse voltage range: 10 V to 1000 V, 0.25 V measurement resolution 
● High-speed sampling testing, fastest 18 ms 
● Contact check function 
● Auto inductance range selection function
● Four-Terminal measurement, ensuring more accurate voltage applied to the inductor 
● Inductance difference voltage compensation function 
● High pulse test sampling rate (200 MHz), 10 bits 
● Breakdown Voltage (BDV) analysis function
● Chinese/English operation interface 
● USB flash drive storage for oscillation waveforms 
● Test settings are automatically saved and recalled on next startup

 

Seven Waveform Judgment Modes

◆  Area Size Comparison

Within any specified A–B interval, the waveform area of the tested coil is integrated and compared with the area of the standard waveform over the same interval. The waveform area is approximately proportional to energy loss, enabling effective detection of interlayer and turn-to-turn short circuits in coils.

◆  Differential Area Comparison

Within any specified A–B interval, the Y-axis difference between the tested coil waveform and the standard waveform is calculated. The resulting difference area is then compared with the standard waveform area over the same interval.

◆  Flutter Value Comparison

Within any specified A–B interval, the first derivative of the waveform is calculated to determine the total discharge energy, which is then compared with that of the standard waveform.

◆  Laplacian Value Comparison

Within any specified A–B interval, the second derivative of the waveform is calculated and compared with that of the standard waveform over the same interval. This method can detect rapid waveform changes caused by electrical discharge or poor electrode soldering.

◆  Peak Ratio Comparison

In Breakdown Voltage (BDV) mode, the peak ratio is calculated using the peak value of the second positive peak versus the first positive peak of the DUT's self-resonant waveform, evaluating whether the product is excessively degraded.

Judgment Formula:

◆  Δ Peak Ratio Comparison

In Impulse Winding Test (IWT) mode, the peak ratio of the DUT's self-resonant waveform is calculated and compared with that of a standard sample. The peak ratio difference reflects variations in energy attenuation caused by parallel resistance (Rp), and can be used to determine whether Rp is abnormal.

Judgment Formula:

◆  Δ Resonant Area Comparison

In IWT mode, after the switch SW is opened, the total area of the self-resonant waveform from the DUT is compared with that of a standard sample. When coil insulation is poor, the resonant waveform decays more rapidly, resulting in a smaller resonant area.

 

Key Technologies

◆  Inductance Auto Range

At the start of testing, the ZX8816 first measures the inductance and automatically selects the appropriate inductance range based on the measured value, ensuring that the DUT is tested under the most suitable waveform conditions.

◆  Contact Check Function

The ZX8816 performs a contact check on the test terminals before high-voltage output, preventing probe arcing caused by poor contact or open circuits. This protects the DUT from damage and extends the service life of the test probes.

◆  Four-Terminal Measurement

Traditional impulse winding testers measure the voltage across both the DUT and the test leads. For low-inductance DUTs, the measured voltage can differ significantly from the actual voltage across the DUT. The ZX8816 employs four-terminal measurement technology to reduce this discrepancy, achieving ideal test results.

◆  Voltage Compensation Function

The ZX8816 provides a voltage compensation function to reduce voltage variations across the inductor caused by inductance differences, thereby minimizing the risk of misjudgment.

 

 

Product Applications

◆  Breakdown Voltage Analysis

The ZX8816 features a breakdown voltage analysis function that allows users to set start voltage, end voltage, and step voltage. During the stepped voltage ramp, the instrument uses three judgment methods—Area Size, Laplacian Value, and Peak Ratio—to compare waveform changes against preset thresholds in real time, thereby determining the dielectric strength of the coil. These functions enable researchers to analyze and optimize the weak points of coil products.

◆  Screenshot Function

Supports screenshot capture. The operator can capture the current test screen at any time via a shortcut key and save it to a USB device in GIF format for easy test documentation.

◆  Data Export Function

Supports data export. Test results can be exported and saved to a USB device with one click in CSV format, facilitating subsequent data analysis and processing.

Parameters & Dimensions

ModelZX8816
Voltage Peak Value10 V to 1000 V
Test Inductance Range0.1 µH  to  1 mH
Voltage Accuracy±(1% * Set Voltage + 2.0 V)
Sampling Rate10bits, 200 MHz
Sampling Ranges8 Ranges: 0,1,2,3,4,5,6,7
Pulse CountTest Pulse Count: 1  to  32; Excitation Pulse Count:0  to  9
Waveform Display Range512*256 points
Detection ModeWaveform Area Size Comparison, Waveform Differential Area Comparison, Waveform Flutter Value Comparison,Waveform Laplacian Value Comparison, Peak Ratio Comparison, Δ Peak Ratio Comparison, and Δ Resonant Area Comparison
Test SpeedFastest 18 ms (Pulse= 1.0)
Sorting IndicationPass/Fail Alarm(Buzzer and LED Indication)
InterfaceRS-232C, HANDLER Interface
Power ConsumptionNo Load:< 150 W; Loaded: < 1000W
Power Supply100  to  240 VAC, 50/60 Hz
Dimensions (W × H × D)430 mm × 177 mm × 449.5 mm
Weight28 kg

 

Unit: millimeters (mm)

Accessories & Options

ZX8816Y01 Test Cable

ZX8816Y01 Test Cable

ZX8816Y03 Test Fixture

ZX8816Y03 Test Fixture

Software & Resources

CLOSE

ZX8816 SMD Power Inductor Testing

ZX8816 vs Chroma19301A Selection Comparison Table

2026-05-11

Solutions

undefined

undefined

Customer Q&A

Q: What are the specific differences between the ZX8816 and the Chroma 19301A across four key dimensions: technical specifications, functional features, application value, and total cost of ownership (TCO)?

A: The ZX8816 and Chroma 19301A are broadly comparable in core specifications. The ZX8816 offers a test speed of 18 ms, while the Chroma typically achieves less than 20 ms — a difference of approximately 2 ms. The ZX8816 provides breakdown point analysis and Pause function, with a local service response time of 4 hours. Overall TCO also differs between the two. For details, please refer to the document "ZX8816 vs Chroma19301A Selection Comparison Table" (PDF). 

(Note: This comparison is compiled from publicly available information from both parties. It does not contain any derogatory statements and is intended solely to save engineers time in conducting their own research, facilitating quick reference for model selection.)

Q: Why are both the waveform and readings abnormal during sampling?

A: This is likely due to incorrect wiring. Please note that the red-to-red test leads should be connected to one end of the inductor, and the black-to-black leads to the other end.

Q: Why is the oscillation amplitude very small during sampling?

A: This may be related to the characteristics of the product. Such products typically require higher power. You can try using the ZX8817 for testing.

Q: Why do the Area or Area Difference values vary significantly during testing?

A: This may be related to the selected start and end points. Generally, the start point should be the first peak on the left, and the end point should be the peak three to four cycles to the right of the start point. Fine-tuning may be applied based on the product characteristics.

Q: Why does the waveform vary significantly during testing, causing frequent misjudgments?

A: This may be due to poor contact at the fixture test terminals. Please carefully check the contact status of the fixture and eliminate factors such as dust contamination.