ZX5545 Series DC Resistance Meter
-
File size: 5.4MB
-
File size: 257.1KB
-
File size: 8.9MB
Product Model
| Model | Overview | Product Brochure |
| ZX5545 | 10-bin sorting, DB37 interface. | Click to View |
| ZX5545-01 | Measurement Range: 0.01 μΩ–1200 MΩ; Temperature measurement function; DB9 sorting interface. | Click to View |
| ZX5545S-10 | 10-channel resistance testing, DB37 interface. | Click to View |
| ZX5545S-20 | 20-channel resistance testing, DB37 interface. | Click to View |
| ZX5545S-02 | Equipped with two scanner boards; 20-channel resistance testing (4-wire) / 42-channel resistance testing (2-wire); DB37 interface. | Click to View |
| ZX5545S-03 | Equipped with one scanner board; 10-channel resistance testing (4-wire) / 21-channel resistance testing (2-wire); DB37 interface. | Click to View |
| ZX5545A | Newly equipped with a 1000 μΩ range; Minimum Resolution: 1nΩ; Maximum Test Current: 1 A; Temperature measurement function; DB37 interface. | Click to View |
Functions & Advantages
● ±0.006% basic accuracy, 0.01 μΩ minimum resolution, 1 A maximum test current ● Measurement range: 0.01 μΩ – 1200 MΩ ● Selectable HIGH/LOW test current for each range | ● Test speed up to 2.2 ms/reading ● Temperature measurement and temperature compensation functions ● Supports RS485 and RS-232C communication protocols for PLC and automation integration | ● OVC (Offset Voltage Compensation) function minimizes the impact of thermal EMF on measurements. ● In LP (Low Power) mode, maximum test voltage can be as low as 20 mV ● Patented USB virtual keyboard technology: data upload to PC without driver or software installation |
Multifunctional Assurance for Measurement Accuracy
◆ Contact Check
Features a comprehensive contact detection function that automatically identifies the contact status of the test fixture. When poor contact is detected, an error signal is output, effectively preventing misjudgment caused by contact issues and eliminating the risk of defective products passing through.

◆ Averaging Function
Averaging from 2 to 16 readings reduces instability, allowing for reliable measurements.
◆ Four-Terminal Measurement Method
Resistance is measured using the four-terminal method, ensuring that the measurement is unaffected by the resistance of the test leads.
◆ Durable Test Probes
The probes feature enhanced bending resistance, and a wide selection of probe options is available to meet diverse measurement needs.

Comparator Function
◆ Audible and visual alarm for Hi/Lo limit exceedance.
◆ Two comparison modes: absolute comparison and relative comparison.
◆ Two distinct audible signals for PASS and FAIL results, with clear differences in frequency, tone, or rhythm for quick identification.
◆ Red and green LED indicators for judgment results, enabling efficient operation without looking at the screen.
Temperature Compensation Function
◆ Temperature Input (Temperature Sensor Port)
Equipped with a temperature sensor port for connecting the dedicated ZX_T001 temperature probe for real-time temperature measurement.
◆ Material- and Temperature-Independent Temperature Compensation
Resistance values that vary with ambient temperature can be dynamically compensated using the built-in algorithm with any user-defined material temperature coefficient, calculating and displaying the reference resistance value at the standard temperature in real time.

◆ Temperature Conversion Function for Temperature Rise Testing
Specifically designed for temperature rise testing, the temperature conversion function directly displays the resistance value at the standard temperature, providing the most critical standardized data for accurate temperature rise calculation and ensuring more reliable results.

Data Logging Function
◆ PC Connection via RS-232C or USB
The instrument can be controlled from a PC to perform various functions and retrieve measurement results. Using patented USB virtual keyboard technology, data can be uploaded directly to the PC without installing USB drivers or PC software.

Multi-Channel Scanner Function
◆ Multi-Point Measurement with Scanner Unit
For applications requiring multi-point measurement—such as network resistors, switching switches, and three-phase motor windings—the scanner unit enables convenient scanning measurements.




Parameters & Dimensions
| Model | ZX5545 |
| Test Parameters | DC Resistance (DCR), Temperature (T) |
| Measurement Range | Normal Mode: 0.00000 mΩ–1200 MΩ |
| LP Mode: 0.00 mΩ–1200 Ω | |
| Range | 10m / 100m / 1000m / 10 / 100 / 1000 / 10k / 100k / 1000k / 10M / 100M / 1000M LP Mode: 1000m / 10 / 100 / 1000 Basic Accuracy: ±0.006% ± 0.001%* of full-scale value |
| Test Current | Normal Mode: 100 nA to 1 A DC LP Mode: 5 μA to 1 mA DC Current Accuracy: ±5% |
| Maximum Open-circuit Voltage | 10 mΩ to 1 kΩ range: 6.5 V 10 kΩ to 1000 MΩ range: 20 V In LP Mode: 20 mV |
| Measurement Method | Four-Terminal measurement method |
| Temperature Compensation | Reference temperature range: -10.0 to 99.9 °C |
| Temperature Coefficient range: -99.999 to 99.999 ppm/°C | |
| Trigger Mode | Internal trigger / External trigger / Bus trigger / Manual trigger |
| Test Speed | Fast (2.2 ms/20 ms) / Medium (40 ms) / Slow (100 ms) |
| Functions | Temperature Compensation / Offset Voltage Compensation / Comparator Sorting / Open‑Circuit Detection / Key Lock |
| Average function | OFF/2–100 times |
| Memory | Internal 50 sets, External USB flash drive 500 sets |
| Interface | RS-232C/RS485, USB DEVICE, USB HOST, HANDLER I/O, foot switch |
| Power Requirements | Voltage: 100 to 120 VAC or 198 to 242 VAC Frequency: 46 to 64 Hz, Power >50 VA |
| Rated power | 40 W |
| Dimensions | 216 mm*87 mm*318 mm |

Unit: millimeters (mm)
Accessories & Options
ZX28Y04 Kelvin Test Cable
ZX-T001 Temperature Probe
Solutions
undefined
Customer Q&A
Q: When connecting to a PLC, the instrument output signal is found to be invalid. How can this be resolved?
A: The Handler interface output mode is set to "Hold" by default. You can change it to "Delay Clear" or "Start Clear" mode.
Q: What should I do if the instrument cannot connect to the PC?
A: ▪ If using a USB connection, check that the bus mode is set to USBHID. Change it in "Setup" → "Bus Mode."
▪ If using an RS-232C connection, verify that the baud rate matches the PC's setting. Adjust it in "Setup" → "Baud Rate."
Q: Why do the measured values differ significantly from the standard values?
A: ▪ Check whether the short-circuit reading (pay attention to the clip connection method) is near 0 mΩ. Re-zero the instrument whenever the test environment changes, such as after replacing the fixture or test leads.
▪ Check whether compensation has been applied. Try disabling compensation and re-zeroing.
▪ If the test leads are long, have multiple adapters, or are aged, inspect all four test leads for open circuits or poor contact.
▪ Verify that the instrument's range matches the DUT. A large range mismatch can cause measurement errors.
Q: Why doesn't the instrument alarm when a defective product is detected?
A: ▪ Make sure the upper and lower limits are set — the instrument only judges after limits are configured.
▪ Check the beep settings. In "Setup", find "Beep" and select either "Pass" or "Fail".
Q: The readings fluctuate significantly and become unstable during zeroing. What should I do?
A: When shorting the clips, ensure that the clips marked with "V" are clamped on the same side. Also, make sure the test lead ports on the instrument are not mismatched — the triangle symbols and colors must correspond correctly.