ZX8817 Series Impulse Winding Tester

The ZX8817 series impulse winding testers are a new generation of pulse insulation testing instruments specifically developed to meet the high-energy requirements of wound components. During the production of coil products (such as transformers, motors, etc.), factors including winding materials, magnetic materials, bobbins, and processing techniques can lead to a degradation in insulation performance between layers, turns, and pins. The ZX8817 stores the sampled waveform of a standard coil in the instrument. During testing, the waveform of the coil under test is compared with the standard waveform, and the quality of the coil is determined based on the configured judgment criteria (Area, Differential Area, Corona, Phase Difference, etc.). This series integrates powerful testing functions, precision sampling and analysis technology, flexible operation methods, and a rich set of interfaces, providing comprehensive testing solutions for most coil products.
  • File size: 334.5KB

  • File size: 5.9MB

  • File size: 1.8MB


Product Model

Model

Overview

Product Brochure

ZX8817-3KVPulse Voltage Output: 300–3000 V; Waveform Sampling Rate: 100 Msps; Minimum Measurable Inductance: 10 mH; Pass/Fail Audible and Visual Alarm Output; Interfaces: Handler, RS-232C, USB Device, USB Host.Click to View
ZX8817-5KVPulse Voltage Output: 300–5000 V; Waveform Sampling Rate: 100 Msps; Minimum Measurable Inductance: 10 mH; Pass/Fail Audible and Visual Alarm Output; Interfaces: Handler, RS-232C, USB Device, USB Host.Click to View
ZX8817A-3KVPulse Voltage Output: 100–3000 V; Waveform Sampling Rate: 200 Msps; Minimum Measurable Inductance: 10 mH; Pass/Fail Audible and Visual Alarm Output; Interfaces: Handler, RS-232C, USB Device, USB Host.Click to View
ZX8817A-5KVPulse Voltage Output: 100–5000 V; Waveform Sampling Rate: 200 Msps; Minimum Measurable Inductance: 10 mH; Pass/Fail Audible and Visual Alarm Output; Interfaces: Handler, RS-232C, USB Device, USB Host.Click to View

Functions & Advantages

● High pulse energy, better suited for wound components with high pulse energy requirements

● 800*480 high-definition large screen display, selectable Chinese/English interface

● Instrument parameters automatically saved, with file loading on startup

● Up to 200 Msps waveform sampling rate, 6500 bytes memory depth

● Firmware upgrade via USB HOST

● Fast test speed with a wide inductance test range

● Four waveform comparison methods: Area, Differential Area, Corona, and Phase Difference

● Excellent voltage stability and test repeatability for consistent results

● Powerful corona analysis and extraction function to identify potential insulation issues

● Waveform zoom and pan functions for detailed observation

● Sample averaging function, supports averaging of up to 32 standard waveforms

● Auto standard acquisition mode automatically selects the appropriate sampling rate

● Degaussing pulse application to eliminate residual magnetism in coils

● Foot switch interface for simple and fast measurements

● Multiple interfaces: Handler, RS-232C, USB Device, USB Host, and RS485

 

Four Comparison Methods

◆  Area Size Comparison 

Within any specified A–B interval, the waveform area of the tested coil is integrated and compared with the area of the standard waveform over the same interval. The percentage difference between the two waveform areas relative to the standard waveform area in that interval serves as the judgment criterion. Since the waveform area is approximately proportional to energy loss, the area comparison method can be used to evaluate energy loss in coils and effectively detect interlayer and turn-to-turn short circuits.

◆  Differential Area Comparison

Within any specified A–B interval, the Y-axis difference between the tested coil waveform and the standard waveform is calculated (the integration result represents the shaded area within the A–B interval). This difference area is compared with the standard waveform area over the same interval, with the judgment baseline set as a percentage. The Differential Area comparison primarily reflects differences in inductance (L) and energy loss, effectively detecting inductance variations between the standard and tested coils.

◆  Corona Discharge Comparison

Independent of waveform differences, within any specified A–B interval, only the high-frequency components contained in the corona discharge spikes of the tested coil waveform are extracted for area (integration) calculation. The result is then compared with a set value (integer) to determine whether the corona discharge level is acceptable.

◆  Phase Difference Comparison

The user can specify a zero-crossing point (from the 2nd to the 10th) as the comparison position. The instrument evaluates the offset between the tested coil waveform and the standard waveform at this zero-crossing point, compares it with the oscillation period of the standard waveform, and uses the percentage difference as the judgment criterion. As shown in the figure, A–B represents the offset, and C–D represents the oscillation period of the standard waveform — here set to compare the third zero-crossing point.

Parameters & Dimensions

ModelZX8817-5KVZX8817-3KVZX8817A-5KVZX8817A-3KV
Output Pulse Voltage

300 V-5000 V

10 V step

5%±10 V

300 V-3000 V

10 V step

5%±10 V

100 V-5000 V

10 V step

5%±10 V

100 V-3000 V

10 V step

5%±10 V

Inductance Test Range>10 μH>10 μH>10 μH>10 μH
Max Pulse EnergyMAX 0.30 JMAX 0.10 JMAX 0.30 JMAX 0.10 J

Display

Waveform Display Area

Display Content

800×480 dot matrix graphic LCD

650×250 dot matrix

Magnified display; setting parameters; standard and test waveforms; comparison results; file information, etc.

Waveform Sampling

Sampling Rate: 100MSa/s to 100kSa/s, 10 ranges;

Resolution: 8 bits;

Sampling Points: 6500 bytes

Sampling Rate: 200MSa/s to 200kSa/s, 10 ranges; 

Resolution: 8 bits; 

Sampling Points: 6500 bytes

Standard Waveform SamplingContinuous Cycle, Single Cycle, Single Sample (Up to 32-sample averaging function)
Input Impedance10 MΩ
Test SpeedUp to 15 meas/sec
Pulse Count

Test pulse: programmable from 1 to 32 cycles

Demagnetization pulses: programmable from 0 to 16 cycles

Comparison Method

AreaSize Comparison, DiffZone Comparison

Corona Comparison, PhaseDiff Comparison

Area Repeatability±1%
Area Difference Repeatability±1%
Judgment OutputPASS/FAIL indication, LED display, Buzzer Alarm
Alarm Volume SelectionLong high beep, Long low beep, Single short beep, Double short beep, OFF
Memory

Internal: 200 sets

External USB flash drive: 500 sets

InterfaceHANDLER, RS-232C, USBCDC, USBTMC, RS485, USBHOST

 

Unit: millimeters (mm)

Accessories & Options

ZX8815Y01B Test Cable

ZX8815Y01B Test Cable

ZX8815Y01R Test Cable

ZX8815Y01R Test Cable

ZX27Y02 Foot Switch

ZX27Y02 Foot Switch

Software & Resources

Solutions

undefined

undefined

Customer Q&A

Q: Why do the Area and Area Difference values fluctuate significantly during testing?

A: ▪ Good product consistency is required; otherwise, significant fluctuation is considered normal.

   ▪ If the product consistency is good and fluctuation still occurs, the selected range may not be optimal. Generally, it is recommended to select the region from the first peak to the third or fourth peak for analysis.

Q: During standard sampling, the waveform appears too sparse or too dense, making it difficult to select a range. How can this be adjusted?

A: You can adjust the waveform density by changing the sampling rate and the time base settings.

Q: Why doesn't the instrument alarm on defects, or why does it keep beeping constantly?

A: Check the beep configuration under "System". Find "Pass Alarm" and "Fail Alarm" and adjust them to your preferred settings.

Q: Why does the instrument display a straight horizontal line during testing?

A: This may be caused by an open circuit resulting in an open-load test, or it could be due to excessive voltage that has damaged the DUT.