ZX6530 Series Insulation Resistance Tester
-
File size: 6.4MB
-
File size: 360.0KB
Product Model
| Model | Overview | Product Brochure |
| ZX6530 | Test Range: 50 kΩ–500 GΩ; Output Voltage: 10 V–500 V (adjustable) | Click to View |
| ZX6533 | Test Range: 50 kΩ–1 TΩ; Output Voltage: 10 V–1000 V (adjustable) | Click to View |
Functions & Advantages
● Test range: 50 kΩ – 500 GΩ / 1 TΩ ● Dual display of insulation resistance and leakage current ● External interfaces: Handler, RS-232C/USB/RS485 ● Test voltage: 10 V – 500 V / 1000 V | ● Fast charge and discharge when measuring capacitive loads ● Automatic data upload via USB interface (patented) ● Equipped with pre-charge voltage interface | ● Front and rear panel test ports simultaneously available ● Freely adjustable charge and measurement time ● Stable negative high-voltage testing |
Negative High-Voltage Testing
DC withstanding voltage testing is a common method for evaluating electrical insulation. By measuring the resistance or leakage current of insulating materials under DC high voltage, potential issues can be identified, allowing for timely maintenance or replacement to ensure reliable insulation under normal operating voltages and reduce the risk of electrical accidents. Compared to AC withstanding voltage testing, DC testing has relatively lower requirements for power supply stability. However, insulation resistance measurement typically requires the use of negative high voltage, primarily for the following reasons.

◆ Easier Detection of Insulation Defects
In actual field operation, the primary factor affecting insulation is moisture ingress, which typically starts from the outer sheath or housing. Water molecules exhibit positive polarity in an electric field.
Positive high voltage repels moisture outward, reducing leakage current and making defects harder to detect. In contrast, negative high voltage attracts moisture toward the high-voltage electrode, significantly increasing leakage current and making defects more apparent.

◆ Reduced External Insulation Flashover
External insulation of equipment typically features highly non-uniform electric fields. When the rod electrode carries a negative charge, the breakdown voltage is significantly higher than with positive polarity.
DC withstanding voltage testing focuses primarily on internal insulation, which is minimally affected by polarity. Therefore, using negative polarity helps prevent external insulation flashover, ensuring safe and effective testing.

◆ Alignment with Actual Operating Conditions
Some electrical equipment may be subjected to negative polarity overvoltages during operation. Testing with negative polarity more closely simulates actual operating conditions, making the evaluation results more valuable for engineering reference.
Equipped with Pre-Charge Voltage Interface
◆ Pre-charge to Achieve Desired Stable Leakage Current
The rear panel of the instrument is equipped with a pre-charge interface, a pre-charge voltage adjustment port, and a secondary charging interface to facilitate pre-charging and achieve the stable leakage current required by the user.
Multiple Advantages for Fast and Accurate Measurement
◆ Nominal Range Mode
Users can manually lock the test range, select auto-ranging, or enable the nominal range mode.

Flexible Adjustment of Test Timing
◆ Charge and Measurement Time
At the moment of voltage application, the DUT generates a transient charging current that is significantly larger than the true insulation leakage current. Immediate measurement at this point would result in serious misjudgment. Therefore, the instrument is designed with an adjustable delay (0–999 s) from charging to the start of measurement, allowing time for the charging current to decay so that measurement is performed only when a stable leakage current is present, ensuring measurement accuracy on automated production lines.

◆ Trigger Delay
To synchronize with the mechanical timing of automated production lines, users can set a delay (0–9999 ms) from the receipt of a trigger signal to the start of the measurement cycle. This ensures that measurement is performed when the DUT is in a stable state, effectively preventing false measurements and misjudgments, and greatly improving the efficiency of automated production line measurements.

Boosting Automated Production Line Efficiency
◆ Front and Rear Panel Test Probe Connections
The instrument can be easily integrated into automated test systems, with test probe connections available on both the front and rear panels, facilitating flexible positioning and angle adjustment in automated production lines.

Parameters & Dimensions
| Model | ZX6530 | ZX6533 |
| Test Parameters | Insulation Resistance (IR) Test | |
| Output Voltage | Adjustable from 10 V to 500 V | Adjustable from 10 V to 1000 V |
| Typical Test Voltage Range | 50 V: 50 kΩ–50 GΩ | |
| 100 V: 100 kΩ–100 GΩ | ||
| 250 V: 250 kΩ–250 GΩ | ||
| 400 V: 400 kΩ–400 GΩ | ||
| 500 V: 500 kΩ–500 GΩ | ||
| \ | 750 V: 750 kΩ–750 GΩ | |
| 1000 V: 1 MΩ–1 TΩ | ||
| Basic Accuracy | In Slow Mode: < 10 GΩ: 3%, ≥ 10 GΩ: 5%, ≥ 100 GΩ: 10% | |
| Pre-Charge Voltage | Synchronized adjusted with test voltage, same range | |
| Max Charging Current | Test Current: 150 mA ± 15 mA | |
| Pre-charge current: 500 mA ± 50 mA | ||
| Range | Auto, Manual, Six-range Test | |
| Test Speed | Fast: 55 meas/sec, Medium: 15 meas/sec, Slow: 3 meas/sec | |
| Correction | Full-range open circuit zeroing | |
| Display Results | Direct reading value | |
| Comparator | Pass, fail, buzzer, LED Display | |
| Interface | RS-232C, RS485, USB, Handler interface | |
| Other | Key lock, data hold function, fixture detection, etc. | |
| Power Requirements | Voltage: 198 to 242 VAC, 50 Hz, Max: 300 VA | Voltage: 198 to 242 VAC, 50 Hz, Max: 500 VA |
| Dimensions (W × H × D) | 400 mm × 132 mm × 415 mm | |
| Weight (kg) | <16 kg | |

Unit: millimeters (mm)
Accessories & Options
ZX65Y01B Leakage Current Test Cable
ZX65Y01R Leakage Current Test Cable
ZX65Y01M Leakage Current Test Cable
ZX26Y04 SMD Test Fixture
Software & Resources
Solutions
undefined
Customer Q&A
Q: After the measurement is completed, the instrument does not discharge automatically and requires a manual trigger. How can this be fixed?
A: Change the test mode from "Single Step" to "Auto" to resolve the issue.
Q: Why can't the instrument connect to the PC?
A: ▪ If using a USB connection, check that the bus mode is set to USBHID. Change it in "System" → "Bus Mode."
▪ If using an RS-232C connection, verify that the baud rate matches the PC's setting. Adjust it in "System" → "Baud Rate."
Q: Why is there a significant deviation between the actual measured value and the standard value when testing a product?
A: ▪ The zeroing may not have been performed properly. Re-zero the instrument whenever the test environment changes, such as after replacing the fixture, changing test leads, or when humidity varies.
▪ If the test leads are long, have multiple adapters, or are aged, check for open circuits or poor contact.
▪ Verify that the instrument's range matches the DUT. A large range mismatch can cause measurement errors. It is generally recommended to use the auto range mode.
Q: Why does the instrument not sound an alarm when a defective product is detected?
A: ▪ Check whether the lower limit has been properly set. The instrument will only perform pass/fail judgment after the limits have been configured.
▪ Verify that the beep settings are enabled. Press and hold the "Settings" key to enter the measurement settings page, locate the "Pass Beep" and "Fail Beep" options, and set them to "ON".
Q: What should I do if the Settings key and System key do not respond when pressed?
A: To prevent accidental parameter changes, both the Settings key and the System key require a long press to activate.