ZX5562/ZX5562A/ZX5562S Precision Battery Tester

The ZX5562/A/S series high-voltage, high-precision battery testers are developed to meet the new demands of the battery industry, specifically for testing low-resistance and large-format lithium battery packs and high-speed sorting on production lines. Leveraging years of technical expertise in impedance testing, Zhixin Precision has enabled the ZX5562 series to achieve a maximum test resolution of 0.1 μΩ and a voltage resolution of 10 μV. The instrument comes standard with RS-232C and HANDLER interfaces, making it easy to set up automated test systems.
  • File size: 4.3MB

  • File size: 472.0KB

  • File size: 7.2MB


Product Model

ModelOverviewProduct Brochure
ZX5562

Resistance Basic Accuracy: ±0.5% rdg. ±5 dgt.; 3 mΩ Range Accuracy: ±0.5% rdg. ±10 dgt.; Voltage Resolution: 10 μV–60 V DC; Resistance Range: 0.1 μΩ–3.1 kΩ; 

Test Frequency: 1 kHz.

Click to View
ZX5562A

Resistance Basic Accuracy: ±0.5% rdg. ±5 dgt.; 3 mΩ Range Accuracy: ±0.5% rdg. ±10 dgt.; Voltage Resolution: 10 μV–300 V DC; Resistance Range: 0.1 μΩ–3.1 kΩ; 

Test Frequency: 1 kHz.

Click to View
ZX5562S-N

Resistance Basic Accuracy: ±0.5% rdg. ±5 dgt.; 3 mΩ Range Accuracy: ±0.5% rdg. ±10 dgt.; Voltage Resolution: 10 μV–60 V DC; Resistance Range: 0.1 μΩ–3.1 kΩ; 

Test Frequency: 1 kHz; N-channel R, N-channel V.

Click to View

 

Functions & Advantages

● 4.3-inch color LCD screen displaying battery internal resistance R and voltage V simultaneously

● Minimum resolution: 0.1 μΩ, 10 μV

● High precision, good stability, fastest test speed up to 100 meas/sec

● Test lead contact check function

● Built-in comparator function, 10-bin sorting

● HANDLER sorting interface for integration with automated equipment

● Standard RS-232C (optional RS485) communication interface, supporting SCPI and Modbus (RTU) protocols

● Test conditions automatically saved: 50 groups internally, 500 setup files on USB flash drive

● Direct upload of test results to PC without software installation

 

Multifunctional Assurance for Measurement Accuracy

◆  Measurement Abnormality Check

Detects poor contact or broken test leads, improving measurement reliability. 

◆  Averaging Function

Averaging from 2 to 16 readings reduces instability, allowing for reliable measurements.

◆  AC Four-Terminal Method

Impedance is measured using the AC four-terminal method, ensuring that the measurement is unaffected by the impedance of the test leads.

 

Comparator Function

◆  Simultaneous impedance and voltage judgment: HI/IN/LO independent judgment for impedance and voltage respectively.

◆  Supports AND logic operation of the two judgment results, outputting PASS/FAIL overall judgment.

◆  Judgment results are indicated via front panel PASS/FAIL LEDs and buzzer, and can also be output through the EXT I/O terminal.

◆  Threshold comparison methods include setting upper/lower limits and setting nominal value with tolerance range.

 

Data Logging Function

◆  PC Application Software

Reads test data to a computer and records it in standard spreadsheet software and CSV files.

 

Accessories

 

Applications

◆  Deterioration assessment and life evaluation of alkaline batteries and lead-acid batteries

◆  UPS testing

◆  Battery factory inspection

◆  High-precision lithium battery internal resistance and voltage testing

◆  High-speed mass production spot checks of coin batteries

◆  Fuel cell stack measurement

◆  Testing of high-voltage battery packs such as lithium batteries in electric vehicles

◆  Internal resistance and voltage measurement of multi-cell series high-voltage lithium batteries

◆  Resistance and voltage measurement of hydrogen fuel cells

◆  Automated battery internal resistance and voltage testing with automatic sorting machines

◆  Battery deterioration status and life evaluation

◆  ESR testing of supercapacitors (low internal resistance capacitors)

◆  Testing of high-voltage battery packs such as lithium batteries in electric vehicles

◆  Internal resistance and voltage measurement of multi-cell series high-voltage lithium batteries

◆  Power battery internal resistance measurement

 

Parameters & Dimensions

Model

ZX5562

ZX5562A

ZX5562S- N

Test Parameters

RV/R/V (R: AC Resistance, V: DC Voltage)

N-channel R, N-channel V

Test Method

Four-Terminal measurement (1 kHz)

Max Input Voltage

Rated input voltage: 60 VDC

Max rated voltage to ground: 60 VDC

Rated input voltage: 300 VDC

Max rated voltage to ground: 300 VDC

Rated input voltage: 60 VDC

Max rated voltage to ground: 60 VDC

Resistance Measurement Range

7 Ranges:

3 mΩ (Max display: 3.1000 mΩ, Resolution: 0.1 μΩ)

30 mΩ (Max display: 31.000 mΩ, Resolution: 1 μΩ)

300 mΩ (Max display: 310.00 mΩ, Resolution: 10 μΩ)

3 Ω (Max display: 3.1000 Ω, Resolution: 0.1 mΩ)

30 Ω (Max display: 31.000 Ω, Resolution: 1 mΩ)

300 Ω (Max display: 310.00 Ω, Resolution: 10 mΩ)

3000 Ω (Max display: 3100.0 Ω, Resolution: 100 mΩ)

Resistance Measurement Accuracy

30 mΩ–3000 Ω Range: ±0.5% rdg. ±5 dgt.
(Add ±3 dgt. for UFAST, add ±2 dgt. for FAST/MED)
 

3 mΩ Range: ±0.5% rdg. ±10 dgt.
(Add ±30 dgt. for UFAST, add ±10 dgt. for FAST, add ±5 dgt. for MED)
 

Voltage Measurement Range

DC 6 V (Resolution: 10 μV)

 DC 60 V (Resolution: 100 μV)

Two ranges switchable

DC 6 V (Resolution: 10 μV)

DC 60 V (Resolution: 100 μV)

DC 300 V (Resolution: 1 mV)

Three ranges switchable

DC 6 V (Resolution: 10 μV)

DC 60 V (Resolution: 100 μV)

Two ranges switchable

Voltage Measurement Accuracy

±0.01% ±3 dgt. (Add ±3 dgt. for UFAST, add ±2 dgt. for FAST, add ±2 dgt. for MED)

Sampling Time

UFAST: 4 ms, FAST: 12 ms, MED: 42 ms, SLOW: 157 ms

(The above is the fastest time for one-parameter measurement at 50 Hz power frequency; depends on test items and power frequency)

Total Measurement Time

Response Time + Sampling Time (Impedance/Voltage response time approx. 1 ms. For reference only; depends on test process and DUT)

Comparator

Independent Judgment: HI/IN/LO (Voltage and impedance judged independently)

Overall Judgment: PASS/FAIL (AND operation of impedance judgment result and voltage judgment result), LED indication, Buzzer alarm, External I/O output

External Interface

Handler I/O, RS-232C, USB

Power Requirements

AC: 85–242 V, 47–63 Hz, 15 VA

Dimensions (W × H × D)

216 mm × 87 mm × 358.5 mm (net dimensions)

226 mm × 104.75 mm × 383.5 mm (with protective cover)

Weight

3.6 kg

 

Unit: millimeters (mm)

Accessories & Options

ZX55Y4549 Four-Wire Alligator Clip

ZX55Y4549 Four-Wire Alligator Clip

ZX55Y4550 Four-Wire Alligator Clip

ZX55Y4550 Four-Wire Alligator Clip

ZX55Y4552 Four-Wire Alligator Clip

ZX55Y4552 Four-Wire Alligator Clip

ZX55Y4553 Four-Wire Resistance Test Probe

ZX55Y4553 Four-Wire Resistance Test Probe

ZX55Y4554 Four-Wire Telescoping Test Rod

ZX55Y4554 Four-Wire Telescoping Test Rod

ZX55Y4555 Four-Wire Telescoping Test Rod

ZX55Y4555 Four-Wire Telescoping Test Rod

ZX55Y4556 Four-Wire Telescoping Battery Test Probe

ZX55Y4556 Four-Wire Telescoping Battery Test Probe

Software & Resources

ZX556X Series PC Software V0.2.5

2026-08-28

ZX556X Series Battery Tester Programming Manual V2.1.5

2026-07-31

Solutions

undefined

undefined

Customer Q&A

Q: What should I do if no value is measured or the readings differ significantly after clamping the clips?

A: Ensure that all four metal jaws of the red and black clips make contact with the two poles of the product. The instrument uses a four-terminal measurement, and each clip jaw operates independently.

Q: After clamping, the measured value slowly rises or falls, or there are slight differences between repeated tests. What is the cause?

A: This is normal. When the internal resistance of the DUT is very small, the influence of contact resistance cannot be ignored, and repeated testing does not guarantee exactly the same clamping position.

Q: Why do the measured values differ significantly from the standard values?

A: ▪ Check whether the short-circuit reading (pay attention to the clip connection method) is near 0 mΩ. Re-zero the instrument whenever the test environment changes, such as after replacing the fixture or test leads.

   ▪ Check whether compensation has been applied. Try disabling compensation and re-zeroing.

   ▪ If the test leads are long, have multiple adapters, or are aged, inspect all four test leads for open circuits or poor contact.

   ▪ Verify that the instrument's range matches the DUT. A large range mismatch can cause measurement errors.

Q: The readings fluctuate significantly and become unstable during zeroing. What should I do?

A: When shorting the clips, ensure that the clips marked with "V" are clamped on the same side. Also, make sure the test lead ports on the instrument are not mismatched — the triangle symbols and colors must correspond correctly.

Total 4 < 1 >

Enter information to retrieve data.

By systematically retrieving and analyzing relevant input information, comprehensive and specialized data are obtained to support or meet specific needs or objectives.

Confirm