Product Model
| Model | Overview | Product Brochure |
| ZX5562 | Resistance Basic Accuracy: ±0.5% rdg. ±5 dgt.; 3 mΩ Range Accuracy: ±0.5% rdg. ±10 dgt.; Voltage Resolution: 10 μV–60 V DC; Resistance Range: 0.1 μΩ–3.1 kΩ; Test Frequency: 1 kHz. | Click to View |
| ZX5562A | Resistance Basic Accuracy: ±0.5% rdg. ±5 dgt.; 3 mΩ Range Accuracy: ±0.5% rdg. ±10 dgt.; Voltage Resolution: 10 μV–300 V DC; Resistance Range: 0.1 μΩ–3.1 kΩ; Test Frequency: 1 kHz. | Click to View |
| ZX5562S-N | Resistance Basic Accuracy: ±0.5% rdg. ±5 dgt.; 3 mΩ Range Accuracy: ±0.5% rdg. ±10 dgt.; Voltage Resolution: 10 μV–60 V DC; Resistance Range: 0.1 μΩ–3.1 kΩ; Test Frequency: 1 kHz; N-channel R, N-channel V. | Click to View |
Functions & Advantages
● 4.3-inch color LCD screen displaying battery internal resistance R and voltage V simultaneously ● Minimum resolution: 0.1 μΩ, 10 μV ● High precision, good stability, fastest test speed up to 100 meas/sec | ● Test lead contact check function ● Built-in comparator function, 10-bin sorting ● HANDLER sorting interface for integration with automated equipment | ● Standard RS-232C (optional RS485) communication interface, supporting SCPI and Modbus (RTU) protocols ● Test conditions automatically saved: 50 groups internally, 500 setup files on USB flash drive ● Direct upload of test results to PC without software installation |
Multifunctional Assurance for Measurement Accuracy
◆ Measurement Abnormality Check
Detects poor contact or broken test leads, improving measurement reliability.

◆ Averaging Function
Averaging from 2 to 16 readings reduces instability, allowing for reliable measurements.
◆ AC Four-Terminal Method
Impedance is measured using the AC four-terminal method, ensuring that the measurement is unaffected by the impedance of the test leads.

Comparator Function
◆ Simultaneous impedance and voltage judgment: HI/IN/LO independent judgment for impedance and voltage respectively.

◆ Supports AND logic operation of the two judgment results, outputting PASS/FAIL overall judgment.
◆ Judgment results are indicated via front panel PASS/FAIL LEDs and buzzer, and can also be output through the EXT I/O terminal.
◆ Threshold comparison methods include setting upper/lower limits and setting nominal value with tolerance range.

Data Logging Function
◆ PC Application Software
Reads test data to a computer and records it in standard spreadsheet software and CSV files.
Accessories

Applications
◆ Deterioration assessment and life evaluation of alkaline batteries and lead-acid batteries
◆ UPS testing
◆ Battery factory inspection
◆ High-precision lithium battery internal resistance and voltage testing
◆ High-speed mass production spot checks of coin batteries
◆ Fuel cell stack measurement
◆ Testing of high-voltage battery packs such as lithium batteries in electric vehicles
◆ Internal resistance and voltage measurement of multi-cell series high-voltage lithium batteries
◆ Resistance and voltage measurement of hydrogen fuel cells
◆ Automated battery internal resistance and voltage testing with automatic sorting machines
◆ Battery deterioration status and life evaluation
◆ ESR testing of supercapacitors (low internal resistance capacitors)
◆ Testing of high-voltage battery packs such as lithium batteries in electric vehicles
◆ Internal resistance and voltage measurement of multi-cell series high-voltage lithium batteries
◆ Power battery internal resistance measurement

Parameters & Dimensions
Model | ZX5562 | ZX5562A | ZX5562S- N |
Test Parameters | RV/R/V (R: AC Resistance, V: DC Voltage) | N-channel R, N-channel V | |
Test Method | Four-Terminal measurement (1 kHz) | ||
Max Input Voltage | Rated input voltage: 60 VDC Max rated voltage to ground: 60 VDC | Rated input voltage: 300 VDC Max rated voltage to ground: 300 VDC | Rated input voltage: 60 VDC Max rated voltage to ground: 60 VDC |
Resistance Measurement Range | 7 Ranges: 3 mΩ (Max display: 3.1000 mΩ, Resolution: 0.1 μΩ) 30 mΩ (Max display: 31.000 mΩ, Resolution: 1 μΩ) 300 mΩ (Max display: 310.00 mΩ, Resolution: 10 μΩ) 3 Ω (Max display: 3.1000 Ω, Resolution: 0.1 mΩ) 30 Ω (Max display: 31.000 Ω, Resolution: 1 mΩ) 300 Ω (Max display: 310.00 Ω, Resolution: 10 mΩ) 3000 Ω (Max display: 3100.0 Ω, Resolution: 100 mΩ) | ||
Resistance Measurement Accuracy | 30 mΩ–3000 Ω Range: ±0.5% rdg. ±5 dgt. 3 mΩ Range: ±0.5% rdg. ±10 dgt. | ||
Voltage Measurement Range | DC 6 V (Resolution: 10 μV) DC 60 V (Resolution: 100 μV) Two ranges switchable | DC 6 V (Resolution: 10 μV) DC 60 V (Resolution: 100 μV) DC 300 V (Resolution: 1 mV) Three ranges switchable | DC 6 V (Resolution: 10 μV) DC 60 V (Resolution: 100 μV) Two ranges switchable |
Voltage Measurement Accuracy | ±0.01% ±3 dgt. (Add ±3 dgt. for UFAST, add ±2 dgt. for FAST, add ±2 dgt. for MED) | ||
Sampling Time | UFAST: 4 ms, FAST: 12 ms, MED: 42 ms, SLOW: 157 ms (The above is the fastest time for one-parameter measurement at 50 Hz power frequency; depends on test items and power frequency) | ||
Total Measurement Time | Response Time + Sampling Time (Impedance/Voltage response time approx. 1 ms. For reference only; depends on test process and DUT) | ||
Comparator | Independent Judgment: HI/IN/LO (Voltage and impedance judged independently) Overall Judgment: PASS/FAIL (AND operation of impedance judgment result and voltage judgment result), LED indication, Buzzer alarm, External I/O output | ||
External Interface | Handler I/O, RS-232C, USB | ||
Power Requirements | AC: 85–242 V, 47–63 Hz, 15 VA | ||
Dimensions (W × H × D) | 216 mm × 87 mm × 358.5 mm (net dimensions) 226 mm × 104.75 mm × 383.5 mm (with protective cover) | ||
Weight | 3.6 kg | ||

Unit: millimeters (mm)
Accessories & Options
ZX55Y4549 Four-Wire Alligator Clip
ZX55Y4550 Four-Wire Alligator Clip
ZX55Y4552 Four-Wire Alligator Clip
ZX55Y4553 Four-Wire Resistance Test Probe
ZX55Y4554 Four-Wire Telescoping Test Rod
ZX55Y4555 Four-Wire Telescoping Test Rod
ZX55Y4556 Four-Wire Telescoping Battery Test Probe
Solutions
undefined
Customer Q&A
Q: What should I do if no value is measured or the readings differ significantly after clamping the clips?
A: Ensure that all four metal jaws of the red and black clips make contact with the two poles of the product. The instrument uses a four-terminal measurement, and each clip jaw operates independently.
Q: After clamping, the measured value slowly rises or falls, or there are slight differences between repeated tests. What is the cause?
A: This is normal. When the internal resistance of the DUT is very small, the influence of contact resistance cannot be ignored, and repeated testing does not guarantee exactly the same clamping position.
Q: Why do the measured values differ significantly from the standard values?
A: ▪ Check whether the short-circuit reading (pay attention to the clip connection method) is near 0 mΩ. Re-zero the instrument whenever the test environment changes, such as after replacing the fixture or test leads.
▪ Check whether compensation has been applied. Try disabling compensation and re-zeroing.
▪ If the test leads are long, have multiple adapters, or are aged, inspect all four test leads for open circuits or poor contact.
▪ Verify that the instrument's range matches the DUT. A large range mismatch can cause measurement errors.
Q: The readings fluctuate significantly and become unstable during zeroing. What should I do?
A: When shorting the clips, ensure that the clips marked with "V" are clamped on the same side. Also, make sure the test lead ports on the instrument are not mismatched — the triangle symbols and colors must correspond correctly.