ZX5544 Series DC Resistance Meter

The ZX5544 series DC resistance meters feature a color LCD display. With a measurement range of 1 μΩ to 3.5 MΩ and a basic accuracy of 0.02%, they are suitable for resistance testing of inductors, chokes, transformers, wire harnesses, fuses, heaters, motors, relays, conductive rubber, and other components. This series is equipped with communication interfaces including RS-232C, RS485, USBTMC, USBCDC, and USBHID, as well as a HANDLER I/O sorting interface, making it particularly suitable for use in automated test equipment.
  • File size: 5.4MB

  • File size: 596.0KB

  • File size: 3.1MB

  • File size: 7.2MB


Product Model

ModelOverviewProduct Brochure
ZX5544Basic Accuracy: ±0.02% ± 5 digits; Measurement Range: 1 μΩ–3.5 MΩ; Temperature measurement function; Minimum Resolution: 1 μΩ; Maximum Test Current: 300 mA; DB37 interface.Click to View
ZX5544-01Basic Accuracy: ±0.02% ± 5 digits; Measurement Range: 1 μΩ–3.5 MΩ; Temperature measurement function; Minimum Resolution: 1 μΩ; Maximum Test Current: 300 mA; DB9 interface.Click to View
ZX5544S-12CHBasic Accuracy: ±0.02% ± 5 digits; Measurement Range: 1 μΩ–3.5 MΩ; Temperature measurement function; 6-channel (12-pin) scanning test; Minimum Resolution: 1 μΩ; Maximum Test Current: 300 mA; DB37 interface.Click to View
ZX5544S-24CHBasic Accuracy: ±0.02% ± 5 digits; Measurement Range: 1 μΩ–3.5 MΩ; Temperature measurement function; 12-channel (24-pin) scanning test; Minimum Resolution: 1 μΩ; Maximum Test Current: 300 mA; DB37 interface.Click to View
ZX5544EX-24CHBasic Accuracy: ±0.02% ± 5 digits; Measurement Range: 1 μΩ–3.5 MΩ; Temperature measurement function; External scanner box, 24-pin scanning test, with arbitrary pin-to-pin combination testing; Minimum Resolution: 1 μΩ; Maximum Test Current: 300 mA; DB37 interface.Click to View
ZX5544EX-32CHBasic Accuracy: ±0.02% ± 5 digits; Measurement Range: 1 μΩ–3.5 MΩ; Temperature measurement function; External scanner box, 32-pin scanning test, with arbitrary pin-to-pin combination testing; Minimum Resolution: 1 μΩ; Maximum Test Current: 300 mA; DB37 interface.Click to View
ZX5544EX-48CHBasic Accuracy: ±0.02% ± 5 digits; Measurement Range: 1 μΩ–3.5 MΩ; Temperature measurement function; External scanner box, 48-pin scanning test, with arbitrary pin-to-pin combination testing; Minimum Resolution: 1 μΩ; Maximum Test Current: 300 mA; DB37 interface.Click to View

Functions & Advantages

● ±0.02% basic accuracy, 1 μΩ minimum resolution, 300 mA maximum test current

● Measurement range: 1 μΩ – 3.5 MΩ

● Test speed up to 4 ms/reading

● Comparator function, OK/NG LED indicators, Handler interface (DB37)

● Key lock function, test conditions automatically saved and recalled on next startup

● Temperature measurement function, customizable temperature compensation

● Optional RS485 interface with MODBUS protocol support for PLC communication

● Patented USB virtual keyboard technology: data upload to PC without driver or software installation

 

Multifunctional Assurance for Measurement Accuracy

◆  Contact Check

Features a comprehensive contact detection function that automatically identifies the contact status of the test fixture. When poor contact is detected, an error signal is output, effectively preventing misjudgment caused by contact issues and eliminating the risk of defective products passing through.

◆  Averaging Function

Averaging from 2 to 16 readings reduces instability, allowing for reliable measurements.

◆  Four-Terminal Measurement Method

Resistance is measured using the four-terminal method, ensuring that the measurement is unaffected by the resistance of the test leads.

◆  Durable Test Probes

The probes feature enhanced bending resistance, and a wide selection of probe options is available to meet diverse measurement needs.

 

Comparator Function

◆  Audible and visual alarm for Hi/Lo limit exceedance.

◆  Two comparison modes: absolute comparison and relative comparison.

◆  Two distinct audible signals for PASS and FAIL results, with clear differences in frequency, tone, or rhythm for quick identification.

◆  Red and green LED indicators for judgment results, enabling efficient operation without looking at the screen.

 

Temperature Compensation Function

◆  Temperature Input (Temperature Sensor Port)

Equipped with a temperature sensor port for connecting the dedicated ZX_T001 temperature probe for real-time temperature measurement.

◆  Material- and Temperature-Independent Temperature Compensation 
Resistance values that vary with ambient temperature can be dynamically compensated using the built-in algorithm with any user-defined material temperature coefficient, calculating and displaying the reference resistance value at the standard temperature in real time.
 

 

◆  Temperature Conversion Function for Temperature Rise Testing

Specifically designed for temperature rise testing, the temperature conversion function directly displays the resistance value at the standard temperature, providing the most critical standardized data for accurate temperature rise calculation and ensuring more reliable results.

 

Data Logging Function

◆  PC Connection via RS-232C or USB

The instrument can be controlled from a PC to perform various functions and retrieve measurement results. Using patented USB virtual keyboard technology, data can be uploaded directly to the PC without installing USB drivers or PC software.

 

Multi-Channel Scanner Function

◆  Multi-Point Measurement with Scanner Unit

For applications requiring multi-point measurement—such as network resistors, switching switches, and three-phase motor windings—the scanner unit enables convenient scanning measurements.

 

Parameters & Dimensions

ModelZX5544ZX5544S-12CHZX5544S-24CHZX5544EX-24CHZX5544EX-32CHZX5544EX-48CH
Test SignalDirect Current (DC)
Test ParametersDC resistance (DCR), temperature (T)
Basic Accuracy0.02% ± 5 digits
Test SpeedUltra-fast: 4 ms Fast: 20 ms Medium: 100 ms Slow: 400 ms
Range ModeAuto, Hold
Trigger ModeInternal, Manual, External, Bus
Test Current300 mA/300 mA/30 mA / 10 mA / 1 mA / 1 mA / 100 μA/5 μA/500 nA
Test Range30 mΩ / 300 mΩ / 3 Ω / 30 Ω / 300 Ω / 3 kΩ / 30 kΩ / 300 kΩ / 3 MΩ
Open-Circuit Port VoltageDC 5.5 V
Temperature Measurement-10 °C – 99.9 °C; Basic Accuracy: ±0.50 °C (combined accuracy when used with the temperature probe ZX-T001) 
Test Speed: 2 s/reading
ComparatorOK, NG (or custom multi‑bin sorting signal)
InterfaceRS-232C, USB DEVICE, USB HOST, HANDLER I/O
ScannerNoneComes with a 12-channel scannerComes with a 24-channel scannerExternal scanner boxExternal scanner boxExternal scanner box
MemoryInternal 50 sets,  External USB flash drive 500 sets
Temperature & Humidity0 °C to -40 °C, Relative Humidity ≤ 75%
Power RequirementsVoltage: 100 to 120 VAC or 198 to 242 VAC
Frequency: 46 to 64 Hz, Power >50 VA
Dimensions (W × H × D)216 mm × 87 mm × 250 mm (net dimensions, without protective cover)
Weight3.2 kg
Attachment1 instruction manual, 1 power cord, 1 test lead

 

 

ZX5544 Series (Single-Channel)

 

 

ZX5544 Series (with Scanning Function)

Unit: millimeters (mm)

Accessories & Options

ZX55Y4549 Four-Wire Alligator Clip

ZX55Y4549 Four-Wire Alligator Clip

ZX-T001 Temperature Probe

ZX-T001 Temperature Probe

ZX55Y4553 Four-Wire Resistance Test Probe

ZX55Y4553 Four-Wire Resistance Test Probe

ZX55Y4554 Four-Wire Telescoping Test Rod

ZX55Y4554 Four-Wire Telescoping Test Rod

ZX55Y4555 Four-Wire Telescoping Test Rod

ZX55Y4555 Four-Wire Telescoping Test Rod

Software & Resources

ZX5544_Zeroing & Temperature Compensation_Operation Demo
CLOSE

ZX5544_Zeroing & Temperature Compensation_Operation Demo

ZX5544_Comparator & Buzzer Settings_Operation Demo
CLOSE

ZX5544_Comparator & Buzzer Settings_Operation Demo

ZX554X Series PC Software V0.2.5

2026-08-28

ZX5544 Series Resistance Tester Programming Manual

2026-07-18

Solutions

undefined

undefined

Customer Q&A

Q: When connecting to a PLC, the instrument output signal is found to be invalid. How can this be resolved?

A: The Handler interface output mode is set to "Hold" by default. You can change it to "Delay Clear" or "Start Clear" mode.

Q: What should I do if the instrument cannot connect to the PC?

A: ▪ If using a USB connection, check that the bus mode is set to USBHID. Change it in "Setup" → "Bus Mode."

   ▪ If using an RS-232C connection, verify that the baud rate matches the PC's setting. Adjust it in "Setup" → "Baud Rate."

Q: Why do the measured values differ significantly from the standard values?

A: ▪ Check whether the short-circuit reading (pay attention to the clip connection method) is near 0 mΩ. Re-zero the instrument whenever the test environment changes, such as after replacing the fixture or test leads.

   ▪ Check whether compensation has been applied. Try disabling compensation and re-zeroing.

   ▪ If the test leads are long, have multiple adapters, or are aged, inspect all four test leads for open circuits or poor contact.

   ▪ Verify that the instrument's range matches the DUT. A large range mismatch can cause measurement errors.

Q: Why doesn't the instrument alarm when a defective product is detected?

A: ▪ Make sure the upper and lower limits are set — the instrument only judges after limits are configured.
   ▪ Check the beep settings. In "Setup", find "Beep" and select either "Pass" or "Fail".

Q: The readings fluctuate significantly and become unstable during zeroing. What should I do?

A: When shorting the clips, ensure that the clips marked with "V" are clamped on the same side. Also, make sure the test lead ports on the instrument are not mismatched — the triangle symbols and colors must correspond correctly.

Total 5 < 1 >