ZX5513 4-Point Probe Sheet Resistance Meter
-
File size: 3.8MB
Product Model
Model | Overview | Product Brochure |
| ZX5513 | Resistance basic accuracy: ±0.05% (resolution 1 μΩ); Range: 20 mΩ–2 MΩ; Supports straight-row and rectangular dual probes; Supports dual geometric correction for thickness and dimensions. |
Functions & Advantages
● Supports simultaneous display of four parameters: R, R□, RY, CY ● Provides "infinite size" and "infinite thickness" modes ● Supports F(W/S) thickness correction and D(d/S) dimension correction ● Nine ranges with automatic matching ● Straight-row or rectangular test probes available | ● Test terminal contact check function ● Highest resistance accuracy: ±0.05%, minimum resolution: 1 μΩ ● Test conditions saved: 50 sets internally, 500 sets on USB flash drive ● Key lock function; automatically saves test conditions on power-off ● Flexible interface functions for customizable data recording, saving, and processing ● Patented technology: results upload to PC without software installation |
Core Measurement Technologies
◆ Sheet Resistance and Resistivity: Core Metrics for Conductivity Assessment
In the R&D and quality inspection of transparent conductive films (such as ITO/FTO), semiconductor epitaxial layers, and metal coatings, sheet resistance (R□) and resistivity (ρ, displayed as RY on the instrument) are the most critical parameters for evaluating whether a thin film's conductivity meets requirements.
Sheet Resistance: Describes "the in-plane resistance of a square thin film." It inherently incorporates thickness information and is the most intuitive conductivity metric for thin-film materials.
Resistivity: A three-dimensional intrinsic parameter of the material that reflects its own conductive capability (determined by carrier concentration and mobility).
The two can be converted via thickness (W), i.e., ρ = R□ × W. This conversion assumes the two-dimensional thin-layer condition, where the film thickness W must be much smaller than the probe spacing S.
◆ DC Four-Point Probe Measurement Principle
During measurement, four equally spaced metal probes contact the sample surface vertically: the two outer probes inject and extract current (I) from a constant current source, while the two inner probes connect to a high-impedance voltmeter to measure voltage (V). Because the voltmeter branch has extremely high impedance, the current flowing through the inner probes is approximately zero, and the contact resistance between the probes and the film produces almost no voltage drop. Therefore, the measured V reflects only the film's own conductivity, fundamentally eliminating contact resistance interference.

By applying a correction factor k that depends on the sample and measurement geometry, the sheet resistance can be calculated: R□ = k × V/I. The entire measurement process requires no ohmic contact electrodes, is non-destructive to the sample, and is fast, forming a complete closed loop of "current excitation → voltage readout → geometric correction → sheet resistance output."
◆ Geometric Correction: Dual Compensation for Size and Thickness
The ideal four-point probe derivation is based on the assumption of "laterally infinite size and vertically extremely thin," i.e., the current field diffuses uniformly from the probes in all directions. In actual measurement, the physical boundaries of the sample truncate the ideal current field, causing internal electric field distortion and deviation of measured values from true values. The instrument must compensate in two independent dimensions:

Dimension Correction (Lateral Boundary Effect): When the sample length and width are comparable to the probe spacing, the current field is compressed at the edges, causing measurement distortion. For small samples, a shape correction factor D(d/S) must be introduced to compensate for electric field distortion caused by edges.
Thickness Correction (Longitudinal 3D Diffusion): The derivation of sheet resistance requires the film thickness to be much smaller than the probe spacing. When the sample is thick, current diffuses three-dimensionally in the depth direction, causing the potential difference measured by the surface probes to be smaller and the apparent sheet resistance to be distorted. A thickness correction factor F(W/S) must then be introduced to compensate for longitudinal current diffusion.
The ZX5513 provides F(W/S) and D(d/S) correction switches and coefficient inputs, enabling reliable measurement results even for non-ideal geometry samples.
Parameters & Dimensions
| Model | ZX5513 |
| Test Parameters | Resistance (R), Sheet Resistance (R□), Resistivity (RY), Conductivity (CY) |
| Measurement Method | DC Four-Point Probe Method (Four-Terminal Measurement) |
| Resolution | 1 μΩ |
| Test Range | Resistance: 1 μΩ to 2 MΩ |
| Measurement Accuracy | Resistance: ±(0.05% + 2 digits) |
| Range Mode | Auto, Lock (Manual) |
| Ranges | 20 mΩ / 200 mΩ / 2 Ω / 20 Ω / 200 Ω / 2 kΩ / 20 kΩ / 200 kΩ / 2 MΩ, 9 ranges total |
| Test Speed | Fast: 40 meas/sec, Medium: 15 meas/sec, Slow: 6 meas/sec |
| Signal Source | Constant current DC 1 μA to 1 A, auto-matched test current per range |
| Probe Type | Straight-row probe, rectangular probe (specified at order) |
| Correction Functions | F(W/S) thickness correction, D(d/S) dimension correction, sample mode (infinite size, infinite thickness) |
| Zeroing Function | Short-circuit zeroing |
| Trigger Mode | Internal (INT), External (EXT) |
| Comparator | Independent upper/lower limit settings for R, R□, RY, CY; PASS/FAIL judgment with beep indication |
| Interfaces | RS-232C, USB Device, USB Host, Handler, GPIB (optional) |
| Power Supply | Voltage: 100 VAC to 120 VAC or 198 VAC to 242 VAC; Frequency: 46 Hz to 64 Hz; Power > 50 VA |
| Dimensions (W × H × D) | 310 mm*110 mm*315 mm |
| Weight (Net) | 4.5 kg |

Unit: millimeters (mm)