ZX1379 Series DC Bias Current Source & LCR Meter with DC Bias Current Source
Category:
-
File size: 5.6MB
-
File size: 451.5KB
Product Model
Model | Overview | Product Brochure |
| ZX1379S | Slave unit; 25 A per unit; Expandable to 225 A with up to 8 units (host required); Frequency Response: 20 Hz–2 MHz. | |
| ZX1379M | Host unit; 25 A per unit; Optional slave units for expansion up to 225 A with up to 8 units; Frequency Response: 20 Hz–2 MHz. | Click to View |
| ZX1379MX | Host unit; 25 A per unit; Optional slave units for expansion up to 225 A with up to 8 units; Frequency Response: 20 Hz–2 MHz; Built-in LCR Meter: 50 Hz–200 kHz, 10 mV–2 V. | Click to View |
| ZX1379AMX | Host unit; 25 A per unit; Optional slave units for expansion up to 225 A with up to 8 units; Frequency Response: 20 Hz–2 MHz; Built-in LCR Meter: 50 Hz–300 kHz, 10 mV–2 V. | Click to View |
| ZX1379BMX | Host unit; 25 A per unit; Optional slave units for expansion up to 225 A with up to 8 units; Frequency Response: 20 Hz–2 MHz; Built-in LCR Meter: 50 Hz–500 kHz, 10 mV–2 V. | Click to View |
| ZX1379CMX | Host unit; 25 A per unit; Optional slave units for expansion up to 225 A with up to 8 units; Frequency Response: 20 Hz–2 MHz; Built-in LCR Meter: 50 Hz–1 MHz, 10 mV–2 V. | Click to View |
Functions & Advantages
● Provides 0 to 25 A constant current output per unit ● Up to 8 slave units, 225 A max constant current output ● Overload alarm indication and output cut-off ● 0.25 mA fine current step | ● 2 MHz frequency response ● Single current and step sweep current output modes ● Graphical operation, Chinese/English interface ● Can be controlled by an external LCR meter via RS-232C interface using SCPI commands |
Main Functions
◆ Wide output range and fine current steps
ZX1373 (0 to 12 A), ZX1376/ZX1379 (0 to 25 A), stackable up to 8 slave units for 225 A total. Minimum step 0.25 mA, dense sampling in the saturation region, accurately constructing inductance vs. bias current curves.
◆ Two current output modes
Single current mode: fixed current, suitable for routine inspection. Step sweep mode: preset current sequence for automatic sweeping, intuitively judging component quality by the rate of inductance decrease.

◆ Bias current output duration
Settable from 0 to 3600 seconds; bias output automatically turns off when the set time is reached (set to 0 for continuous output). Prevents device overheating, suitable for production lines with strict control over test duration.
◆ Auto current continuation
When changing DUTs, the current is automatically adjusted to near 0 A without turning off the bias; it restores once the DUT is re-clamped. No manual start/stop required, improving mass testing efficiency.
◆ Open/short correction
To eliminate parasitic parameter errors introduced by test fixtures and connecting cables, ensuring measurement results truly reflect DUT characteristics.
Typical Applications
◆ Current saturation performance testing
Using the "Changzhou ZhiXin Inductor DC Bias Characteristic Analysis PC Software," measurement data and curve graphics are displayed intuitively, generating inductance vs. bias current characteristic curves and accurately calibrating the saturation limited current.

◆ Temperature rise current testing
Based on the collaborative operation of the inductor superimposed current source and the temperature tester, the dynamic characteristic curve of inductor temperature rise versus current can be collected and recorded in real time, accurately presenting the correspondence between temperature and current, thereby accurately calibrating the temperature rise limited current.

Parameters & Dimensions
| Model | ZX1379M | ZX1379MX | ZX1379S | |
| Display | 7" LCD | 7" LCD | LED Indicator Light | |
| Operation | Physical Buttons + Foot Switch | Physical Buttons + Foot Switch | Controlled by the host | |
| Supports Measurement Frequency | up to 2 MHz | |||
| Current Range | 0–25 A | 0–25 A | 25 A per unit; Up to 8 units stackable for 225 A total | |
| Current | Range | 0 mA to 1.000 A | 1.000 A to 5.000 A | 5.0 A to 225 A |
| Step | 0.25 mA | 1.25 mA | 6.25 mA × N (number of online units) | |
| Scan Adjustment Time | 4 ms to 3600 s | 10 ms to 3600 s | 20 ms to 3600 s | |
| Accuracy | ±(1% + 5 mA) | ±2% | ±3% | |
| Max Output Voltage | 11 V (Can output 11 V under all current conditions!!!) | |||
| Max Allowed DC Resistance | Rmax = Vmax / I (Ω) (Refer to the user manual for Rmax calculation details) | |||
| Max Allowed Inductance | Lmax = Vmax / (di/dt) (mH) (Refer to the user manual for Lmax calculation details) | |||
| Range Mode | Auto | |||
| Start/Stop Control Mode | START/STOP Physical Button, 4 Type of Foot Controls, Bus | |||
| Max Continuous Current Load Time | Conservative 2–3 hours, uninterrupted output | |||
| Functions | Instrument fault self-test; 99 groups of custom file management; Chinese/English bilingual; SCPI command set | |||
| Interface | RS-232C, SlaverLink | SlaverLink | ||
| Dimensions (W × H × D) | 433.5 mm × 177 mm × 464.2 mm | |||
| Weight (net weight) | 19 kg | |||
| Model | ZX1379MX | ZX1379AMX | ZX1379BMX | ZX1379CMX | |
| Test Frequency | 50 Hz to 200 kHz | 50 Hz to 300 kHz | 50 Hz to 500 kHz | 50 Hz to 1 MHz | |
| More than 12,000 frequency points | |||||
| Test Parameters | C, L, X, B, R, G, D, Q, ESR, Rp, θ, |Z|, |Y| | ||||
| Test Level | 10 mV-2 Vrms | ||||
| Basic Accuracy | 0.05%, refer to accuracy table | ||||
| Test Speed | Fast: 50 meas/sec, Medium: 20 meas/sec, Slow: 6 meas/sec (Frequencies above 1 kHz) | ||||
| Display Range | |Z|, R, X | 0.01 mΩ to 99.9999 MΩ | |||
| |Y|, G, B | 0.00001 μS to 99.9999 S | ||||
| C | 0.00001 pF to 9.99999 F | ||||
| D | 0.00001 to 9.99999 | ||||
| L | 0.00001 μH to 99999.9 H | ||||
| Q | 0.01 to 99999.9 | ||||
| θ(RAD) | -3.14159 to 3.14159 | ||||
| θ(DEG) | -179.999° to 179.999° | ||||
| Δ% | -99.999% to 99.999% | ||||
| Signal Source Output Impedance | 30 Ω, 100 Ω, 1062X | ||||
| Range Mode | Auto, Hold | ||||
| Trigger Mode | Internal, Manual, Auto, External, Bus | ||||
| Correction Function | Open/Short, All-frequency zeroing, Spot-frequency zeroing | ||||
| Equivalent Circuit | Series, Parallel | ||||
| Comparator | 10-bin sorting, BIN0–BIN9 | ||||
| PASS, HI, LOW (PASS/FAIL LED indication) | |||||

Unit: millimeters (mm)
Accessories & Options
ZX1375Y02 Maximum 50 A Superimposed Current Test Cable
ZX1379Y05 Maximum Superimposed Current 250 A Test Fixture
ZX27Y01 DB25 Transformer Scan Box Control Cable
ZX1378SY01 Slave Current Expansion Bar
ZX28Y11 Four-Terminal Kelvin Test Cable
ZX1379Y07 SMD High-Current Test Fixture
ZX1378Y03B Large-Current Test Clip for Bias Current Source
ZX1378Y03R Large-Current Test Clip for Bias Current Source
Software & Resources
Solutions
undefined
Customer Q&A
Q: Why does the instrument display a "BIAS OVER" error?
A: This is typically caused by an open circuit in the test leads, preventing the loop from being properly formed. It may also be due to the internal resistance of the DUT being too high.
Q: Why does the instrument fail to alarm when a defective product is detected, or why does it keep alarming continuously?
A: ▪ Check whether the upper/lower limits and nominal value have been set, and ensure the comparator function is enabled.
▪
Q: The reading fluctuates significantly when the instrument applies bias current. How can this be resolved?
A: ▪ Check whether the test leads are making reliable contact with the DUT.
▪ Check the instrument's internal resistance setting. Try changing it to 30 Ω or refer to the 1062X series for comparison.
Q: Why do the readings fluctuate significantly and become unstable during zeroing?
A: When shorting the clips, ensure that the corresponding color-matched clips (red-to-blue and black-to-white) are clipped on the same side.
Q: When connecting to a PLC, the instrument output signal is found to be invalid. How can this be resolved?
A: The Handler interface output mode is set to "Hold" by default. You can change it to "Delay Clear" or "Start Clear" mode.
Q: Why do the measured values differ significantly from the standard values?
A: ▪ Check whether the short-circuit reading (pay attention to the clip connection method) is near 0 mΩ. Re-zero the instrument whenever the test environment changes, such as after replacing the fixture or test leads.
▪ Check whether compensation has been applied. Try disabling compensation and re-zeroing.
▪ If the test leads are long, have multiple adapters, or are aged, inspect all four test leads for open circuits or poor contact.
▪ Verify that the instrument's range matches the DUT. A large range mismatch can cause measurement errors.