ZX6583 Series Insulation Resistance Tester
-
File size: 261.1KB
-
File size: 5.7MB
-
File size: 586.6KB
Product Model
Model | Overview | Product Brochure |
| ZX6583 | Output Voltage: 1 V–650 V; IR Measurement Range: 100 kΩ–5 TΩ; Charging current: 200 mA. | |
| ZX6583A | Output Voltage: 1 V–1000 V; IR Measurement Range: 100 kΩ–10 TΩ; Charging current: 100 mA. |
Functions & Advantages
● Color LCD display ● Chinese/English operation interface ● Measurement range: 1 nA–20 mA ● Auto or manual ranging ● Open zeroing function | ● Contact check function for capacitive materials ● SCPI commands for easier integration into automated systems ● Test conditions automatically saved; supports USB flash drive file storage ● Comparator function for pass/fail sorting ● Instrument software upgrade via USB flash drive |
Negative High-Voltage Testing
DC withstanding voltage testing is a common method for evaluating electrical insulation. By measuring the resistance or leakage current of insulating materials under DC high voltage, potential issues can be identified, allowing for timely maintenance or replacement to ensure reliable insulation under normal operating voltages and reduce the risk of electrical accidents. Compared to AC withstanding voltage testing, DC testing has relatively lower requirements for power supply stability. However, insulation resistance measurement typically requires the use of negative high voltage, primarily for the following reasons.

◆ Easier Detection of Insulation Defects
In actual field operation, the primary factor affecting insulation is moisture ingress, which typically starts from the outer sheath or housing. Water molecules exhibit positive polarity in an electric field.
Positive high voltage repels moisture outward, reducing leakage current and making defects harder to detect. In contrast, negative high voltage attracts moisture toward the high-voltage electrode, significantly increasing leakage current and making defects more apparent.

◆ Reduced External Insulation Flashover
External insulation of equipment typically features highly non-uniform electric fields. When the rod electrode carries a negative charge, the breakdown voltage is significantly higher than with positive polarity.
DC withstanding voltage testing focuses primarily on internal insulation, which is minimally affected by polarity. Therefore, using negative polarity helps prevent external insulation flashover, ensuring safe and effective testing.

◆ Alignment with Actual Operating Conditions
Some electrical equipment may be subjected to negative polarity overvoltages during operation. Testing with negative polarity more closely simulates actual operating conditions, making the evaluation results more valuable for engineering reference.
Flexible Adjustment of Test Timing
◆ Charge and Measurement Time
At the moment of voltage application, the DUT generates a transient charging current that is significantly larger than the true insulation leakage current. Immediate measurement at this point would result in serious misjudgment. Therefore, the instrument is designed with an adjustable delay (0–999 s) from charging to the start of measurement, allowing time for the charging current to decay so that measurement is performed only when a stable leakage current is present, ensuring measurement accuracy on automated production lines.

◆ Trigger Delay
To synchronize with the mechanical timing of automated production lines, users can set a delay (0–9999 ms) from the receipt of a trigger signal to the start of the measurement cycle. This ensures that measurement is performed when the DUT is in a stable state, effectively preventing false measurements and misjudgments, and greatly improving the efficiency of automated production line measurements.

Boosting Automated Production Line Efficiency
◆ Front and Rear Panel Test Probe Connections
The instrument can be easily integrated into automated test systems, with test probe connections available on both the front and rear panels, facilitating flexible positioning and angle adjustment in automated production lines.

Parameters & Dimensions
| Model | ZX6583 | ZX6583A |
| Test Parameters | Insulation resistance IR, leakage current LC | Insulation resistance IR, leakage current LC |
| Output Voltage | 1 V to 650 V | 1 V to 1000 V |
| Voltage Accuracy | When Voltage ≥ 10 V: 1% ± 1 V | When Voltage ≥ 10 V: 1% ± 1 V |
| When Voltage <10 V: 10% ± 0.1 V | When Voltage <10 V: 10% ± 0.1 V | |
| Resistance Measurement Range | 100 kΩ to 5 TΩ | 100 kΩ to 10 TΩ |
| Resistance Measurement Accuracy | IR < 1 MΩ: 5%, IR ≥ 1 MΩ: 1%, IR ≥ 1 GΩ: 3%, IR ≥ 10 GΩ: 5%, IR ≥ 1 TΩ: 10% | |
| Range | Auto, Manual | Auto, Manual |
| Charging Current | 200 mA (MAX) | 100 mA(MAX) |
| Test Speed | 55 meas/sec (fast), 15 meas/sec (slow) | 55 meas/sec (fast), 15 meas/sec (slow) |
| Display Results | Current, Resistance, Sorting results, Audio‑Visual Alarm | Current, Resistance, Sorting results, Audio‑Visual Alarm |
| Charging Timer | 999.9 s | 999.9 s |
| Interface | RS-232C interface, USB, HANDLER Interface | RS-232C interface, USB, HANDLER Interface |
| General Characteristics | ||
| Power Requirements | Voltage: 198 V to 240 VAC Frequency: 50 Hz Power: Max 75 VA | Voltage: 198 V to 240 VAC Frequency: 50 Hz Power: Max 75 VA |
| Dimensions (W × H × D) | 310 mm × 110 mm × 323.5 mm | 310 mm × 110 mm × 323.5 mm |
| Weight | 6 kg | 6 kg |

Unit: millimeters (mm)
Accessories & Options
ZX65Y01B Leakage Current Test Cable
ZX65Y01R Leakage Current Test Cable
ZX65Y01M Leakage Current Test Cable
ZX65Y04 SMD Test Fixture
Software & Resources
Solutions
undefined
Customer Q&A
Q: During testing, the insulation resistance value is too low. How can this be improved?
A: A small current exists in the loop. Perform open-circuit zeroing (zeroing test) during no-load measurement to clear the current offset, which will improve the result and make the data accurate.
Q: The instrument does not make a judgment, and the buzzer does not sound. How should this be handled?
A: The upper/lower limits of insulation resistance must be set for the instrument to make a judgment. The buzzer sound must be set in the system settings by locating "Pass Beep" and "Fail Beep."
Q: During testing, the insulation resistance value keeps changing. What is the cause?
A: Some products are not pure resistors and have a capacitive component; this is normal. To avoid it, pre-charge the product.