Product Categories
ZX9310/ZX9320/ZX9320S Series Hipot Tester
-
File size: 454.5KB
-
File size: 2.1MB
Product Model
Model | Overview | Product Brochure |
| ZX9310 | Hipot Tester (AC/DC/IR); Max AC Current: 10 mA; Voltage: AC 0.05 kV–5.00 kV ±2%, DC 0.05–6.00 kV ±2%. | Click to View |
| ZX9310A | Hipot Tester (AC/DC); Max AC Current: 10 mA; Voltage: AC 0.05 kV–5.00 kV ±2%, DC 0.05–6.00 kV ±2%. | Click to View |
| ZX9310B | Hipot Tester (AC); Max AC Current: 10 mA; Voltage: AC 0.05 kV–5.00 kV ±2%. | Click to View |
| ZX9320 | Hipot Tester (AC/DC/IR); Max AC Current: 20 mA; Voltage: AC 0.05 kV–5.00 kV ±2%, DC 0.05–6.00 kV ±2%; IR: 50–1000 V ±2%. | Click to View |
| ZX9320A | Hipot Tester (AC/DC); Max AC Current: 20 mA; Voltage: AC 0.05 kV–5.00 kV ±2%, DC 0.05–6.00 kV ±2%. | Click to View |
| ZX9320B | Hipot Tester (AC); Max AC Current: 20 mA; Voltage: AC 0.05 kV–5.00 kV ±2%. | Click to View |
| ZX9320S-2 | Hipot Tester (AC/DC/IR); Max AC Current: 20 mA; With 2-channel scanning interface; Voltage: AC 0.05 kV–5.00 kV ±2%, DC 0.05–6.00 kV ±2%. | Click to View |
| ZX9320S-4 | Hipot Tester (AC/DC/IR); Max AC Current: 20 mA; With 4-channel scanning interface; Voltage: AC 0.05 kV–5.00 kV ±2%, DC 0.05–6.00 kV ±2%. | Click to View |
| ZX9320S-8 | Hipot Tester (AC/DC/IR); Max AC Current: 20 mA; With 8-channel scanning interface; Voltage: AC 0.05 kV–5.00 kV ±2%, DC 0.05–6.00 kV ±2%. | Click to View |
Functions & Advantages
| ● 4.3" TFT-LCD for complete test result display, improving reading efficiency ● List sweep display for one-screen viewing of multi-step test results ● Key lock function protects memory data; automatically saves test conditions on power-off | ● Stores up to 100 test steps, with 4 test modes that can be freely combined ● Current baseline zeroing function | ● Electric wall (GFI) technology continuously monitors the circuit to prevent electric shock and ensure operator safety (compatible with ZX9501) ● Voltage rise time, test time, and fall time can be independently set within 999.9 seconds; insulation resistance wait time is also fully adjustable ● PLC interface for easy integration into automated test systems |
Stable Output Unaffected by AC Power Fluctuations
Equipped with closed-loop feedback and linear amplifier technology for precise output voltage control — stable even under load variations!
Traditional hipot testers typically use conversion transformers for direct step-up, making the output voltage susceptible to AC line input fluctuations, which may lead to inaccurate test results. When distorted voltage is applied to the DUT, it may accelerate component degradation and create hidden quality issues. The ZX9320 series employs closed-loop output voltage control, monitoring and adjusting the output in real time to ensure voltage fluctuation within ±(1% reading + 10 V) under various DUT load conditions. The internal independent regulated power supply design effectively reduces the impact of external grid disturbances, providing a solid foundation for testing.

High Precision and High-Speed Measurement
The ZX9320 series delivers high-precision measurement capability with voltage measurement accuracy of ±2.0% and current measurement accuracy of ±(5.0% reading + 5 digits). Equipped with auto-ranging, it achieves high-speed judgment with a minimum test time of 0.2 seconds, ensuring fast and accurate measurements.
◆ Current Zeroing Function
During testing, changes in ambient temperature and humidity as well as different cable routing can cause stray currents from distributed capacitance in the test leads and fixtures, resulting in a baseline reading under open-circuit conditions. The current zeroing function eliminates such stray current interference with one-touch, further improving measurement accuracy and ensuring reliable insulation performance evaluation.

◆ Arc Detection Function
Specifically designed to capture transient discharge pulses in the high-voltage test loop, filtering out normal current and responding only to high-speed current variations. Due to low-pass filter characteristics and the inherent randomness of arc magnitude, this function can only provide an approximate assessment of localized arcing severity.

Precise Time Control of Applied Voltage
Features wide-range voltage rise/fall time control! Prevents adverse effects of over-spec voltage surges on the DUT.
The ZX9320 series supports both rise time and fall time control in AC withstand, DC withstand, and insulation resistance tests. Unlike traditional instruments that only allow short fixed rise/fall times, the ZX9320 series allows rise or fall times to be set up to 999.9 seconds. This capability is especially useful for large capacitive loads or DUTs sensitive to voltage transients, effectively preventing insulation damage or component aging caused by transient overvoltage and ensuring tests are both rigorous and safe.
◆ Rise Time Control Function
Users can set the time for voltage to rise from 0 to the set value based on DUT characteristics, adjustable from 0.1 s to 999.9 s (can be disabled).

◆ Fall Time Control Function
After the test, users can set the time for voltage to smoothly decrease from the set value to 0, adjustable from 0.1 s to 999.9 s (can be disabled).

Active Safety Protection Mechanisms
◆ Interlock Switch (INTERLOCK)
INTERLOCK is used for multi-unit interconnection, preventing damage caused by simultaneous output from multiple instruments due to misoperation. When the interlock switch is cut off, the instrument forcibly inhibits high-voltage startup. In other words, the interlock loop being closed is a prerequisite for the instrument to output high voltage.
◆ Forced Fast Discharge Function
After DC withstand or insulation resistance testing, the instrument automatically discharges high-voltage charge stored in capacitive loads to a safe voltage level, preventing residual electric shock hazards.
◆ Open/Short Circuit Detection (OSC)
Pre-tests the test loop before high-voltage application, capable of detecting distributed impedance currents above 100 pF to confirm reliable connections, preventing misjudgment caused by open test leads or poor pin contact, and ensuring accurate and safe high-voltage testing.
Multi-Step Results on One Screen
◆ List Sweep Function
Users can freely edit multiple measurement steps based on test requirements, with each step independently configurable for test voltage, test time, upper/lower limits, and other parameters. Enables one-screen display of measurement results from different steps and conditions.

PC Communication Software
The companion software communicates with the instrument via RS-232C, supporting parameter setting, test initiation, and data monitoring. Test results can be exported to Excel format for quality traceability and data analysis.

Wide Range of Applications
◆ Power Equipment and Grid Maintenance
◆ Cable and Transformer Manufacturing
◆ New Energy Sector
◆ Industrial Electrical Equipment
◆ Rail Transit and Aviation
◆ Laboratories and Certification Bodies

Parameters & Dimensions
| Model | ZX9310 | ZX9310A | ZX9310B | ZX9320 | ZX9320A | ZX9320B | |
| Withstanding Voltage Test | |||||||
| Output Voltage | AC | 0.05 kV to 5.00 kV ±2% | |||||
| DC | 0.05 to 6.00 kV ±2% | 0.05 to 6.00 kV ±2% | ------ | 0.05 to 6.00 kV ±2% | 0.05 to 6.00 kV ±2% | ------ | |
| Voltage Adjustment Rate | ≤(1.0%+10 V)(Rated Power) | ||||||
| Current Test Range | AC | 0 to 10.00 mA | 0 to 10.00 mA | 0 to 10.00 mA | 0 to 20.00 mA | 0 to 20.00 mA | 0 to 20.00 mA |
| DC | 0 to 5.00 mA | 0 to 5.00 mA | ------ | 0 to 10.00 mA | 0 to 10.00 mA | ------ | |
| Test Accuracy | ±(5.0% Reading+5 digits) | ||||||
| Fast Discharge | Automatic discharge after test (DCW) | ------ | Automatic discharge after test (DCW) | ------ | |||
| Insulation Resistance Test | |||||||
| Output Voltage | 0.05 to 1.00 kV ±2% | ------ | ------ | 0.05 to 1.00 kV ±2% | ------ | ------ | |
| Resistance Test Range | 0.1 to 50 GΩ | ------ | ------ | 0.1 to 50 GΩ | ------ | ------ | |
Resistance Test Accuracy | 50 V to 1000 V | 0.1 to 999 MΩ ±10% | ------ | ------ | 0.1 to 999 MΩ ±10% | ------ | |
| 1.00 to 10 GΩ ±20% | ------ | ------ | 1.00 to 10 GΩ ±20% | ------ | ------ | ||
| 500 V to 1000 V | 10 GΩ to 50 GΩ (25% of reading) | ------ | ------ | 10 GΩ to 50 GΩ (25% of reading) | ------ | ------ | |
| Discharge Function | Auto Discharge After Test(DCW) | ------ | ------ | Auto Discharge After Test(DCW) | ------ | ------ | |
| Arc Detection | |||||||
| Measurement Range | AC | Levels 1 to 9(Factory default is 5) (Corresponding current: 20 mA, 18 mA, 16 mA,14 mA, 12 mA, 10 mA,7.7 mA, 5.5 mA, 2.8 mA) | |||||
| DC | Levels 1 to 9 | ------ | Levels1 to 9 | ------ | |||
| General Parameters | |||||||
| Memory | 120 sets, 100 steps per set | ||||||
| Voltage Rise Time | 0.1 s to 999.9 s | ||||||
| Voltage Fall Time | 0.1 s to 999.9 s | ||||||
| Voltage Wait Time | 0.2 s to 999.9 s | 0.2 s to 999.9 s | |||||
| Test Time Setting | 0.2 s to 999.9 s | ||||||
| Interface | |||||||
| Standard | HANDLER, RS-232C, USB HOST (USB Flash Drive Interface) | ||||||
| Dimensions (W × H × D) | 310x110x415 mm | ||||||
| Weight | 15.5 kg | ||||||

Unit: millimeters (mm)
Accessories & Options
ZX95Y02B Test Cable
ZX95Y02R Test Cable
ZX95Y01R High-Voltage Test Probe
ZX95Y03B High-Voltage Test Rod
ZX95Y03R High-Voltage Test Rod
Solutions
undefined
Customer Q&A