ZX8815/ZX8813/ZX8812 Series Impulse Winding Tester
Category:
-
File size: 445.5KB
-
File size: 5.5MB
-
File size: 2.3MB
Product Model
| Model | Overview | Product Brochure |
| ZX8812-2KV | Pulse Voltage Output: 50 V–2000 V; Waveform Sampling Rate: 100 Msps. | Click to View |
| ZX8812 A-2KV | Pulse Voltage Output: 50 V–2000 V; Waveform Sampling Rate: 200 Msps. | Click to View |
| ZX8813 | Pulse Voltage Output: 30 V–1000 V; Waveform Sampling Rate: 100 Msps; Micro-inductance testing. | Click to View |
| ZX8813A | Pulse Voltage Output: 30 V–1000 V; Waveform Sampling Rate: 200 Msps; Micro-inductance testing. | Click to View |
| ZX8815-3KV | Pulse Voltage Output: 100 V–5000 V; Waveform Sampling Rate: 100 Msps. | Click to View |
| ZX8815A-3KV | Pulse Voltage Output: 100 V–3000 V; Waveform Sampling Rate: 200 Msps. | Click to View |
| ZX8815-5KV | Pulse Voltage Output: 100 V–5000 V; Waveform Sampling Rate: 100 Msps. | Click to View |
| ZX8815A-5KV | Pulse Voltage Output: 100 V–5000 V; Waveform Sampling Rate: 200 Msps. | Click to View |
Functions & Advantages
● 65K-color 7-inch TFT 800*480 HD display, selectable Chinese/English interface ● 100 Msps waveform sampling rate, 6500 bytes memory depth (customizable to 200 Msps) ● High-speed testing, up to 15 meas/sec ● Four waveform comparison methods: Area, Differential Area, Corona, and Phase Difference ● Excellent voltage stability and test repeatability for consistent results | ● Powerful corona analysis and extraction function to identify potential insulation issues ● Instrument parameters automatically saved with file loading on startup ● Waveform vertical zoom, horizontal zoom, and pan functions for detailed observation ● Sample averaging function, supporting averaging of up to 32 standard waveforms ● Auto standard acquisition mode automatically selects the appropriate sampling rate | ● Destructive testing mode to help select the appropriate test voltage ● Degaussing pulse application ensures waveform consistency ● Firmware upgrade via USB HOST ● Foot switch interface for simple and fast measurements ● Multiple interfaces: Handler, RS-232C, USB Device, USB Host, and GPIB (optional) |
Four Comparison Methods
◆ Area Comparison
Within any specified A–B interval, the waveform area of the tested coil is integrated and compared with the area of the standard waveform over the same interval. The percentage difference between the two waveform areas relative to the standard waveform area in that interval serves as the judgment criterion. Since the waveform area is approximately proportional to energy loss, the area comparison method can be used to evaluate energy loss in coils and effectively detect interlayer and turn-to-turn short circuits.

◆ Differential Area Comparison
Within any specified A–B interval, the Y-axis difference between the tested coil waveform and the standard waveform is calculated (the integration result represents the shaded area within the A–B interval). This difference area is compared with the standard waveform area over the same interval, with the judgment baseline set as a percentage. The Differential Area comparison primarily reflects differences in inductance (L) and energy loss, effectively detecting inductance variations between the standard and tested coils.

◆ Corona Discharge Comparison
Independent of waveform differences, within any specified A–B interval, only the high-frequency components contained in the corona discharge spikes of the tested coil waveform are extracted for area (integration) calculation. The result is then compared with a set value (integer) to determine whether the corona discharge level is acceptable.

◆ Phase Difference Comparison
The user can specify a zero-crossing point (from the 2nd to the 10th) as the comparison position. The instrument evaluates the offset between the tested coil waveform and the standard waveform at this zero-crossing point, compares it with the oscillation period of the standard waveform, and uses the percentage difference as the judgment criterion. As shown in the figure, A–B represents the offset, and C–D represents the oscillation period of the standard waveform — here set to compare the third zero-crossing point.

Parameters & Dimensions
| Model | ZX8813 | ZX8812-2KV | ZX8813A | ZX8812 A-2KV |
| Output Pulse Voltage | 30 V-1000 V 10 V step 5%±10 V | 50 V-2000 V 10 V step 5%±10 V | 30 V-1000 V 10 V step 5%±10 V | 50 V-2000 V 10 V step 5%±10 V |
| Inductance Test Range | >1 μH | >5μH | >1 μH | >5μH |
| Max Pulse Energy | MAX 0.02 J | MAX 0.08 J | MAX 0.02 J | MAX 0.08 J |
Display Waveform Display Area Display Content | 800×480 dot matrix graphic LCD 650×250 dot matrix Magnified display; setting parameters; standard and test waveforms; comparison results; file information, etc. | |||
| Waveform Sampling | Sampling Rate: 100MSa/s to 100kSa/s, 10 ranges Resolution: 8 bits Sampling Points: 6500 bytes | Sampling Rate: 200MSa/s to 200kSa/s, 10 ranges Resolution: 8 bits Sampling Points: 6500 bytes | ||
| Standard Waveform Sampling | Continuous Cycle, Single Cycle, Single Sample (Up to 32-sample averaging function) | |||
| Input Impedance | 10 MΩ | |||
| Test Speed | Up to 15 meas/sec | |||
| Pulse Count | Test pulse: programmable from 1 to 32 cycles Demagnetization pulses: programmable from 0 to 16 cycles | |||
| Display Waveform Measurement | Voltage, Time, Frequency | |||
| Trigger Mode | Internal / Manual (including foot control) / External / Bus | |||
| Comparison Method | AreaSize Comparison, DiffZone Comparison Corona Comparison, PhaseDiff Comparison | |||
| Area Repeatability | ±1% | |||
| Area Difference Repeatability | ±1% | |||
| Judgment Output | PASS/FAIL indication, LED display, Buzzer Alarm | |||
| Alarm Volume Selection | Long high beep, Long low beep, Single short beep, Double short beep, OFF | |||
| Memory | Internal: 200 sets External USB flash drive: 500 sets | |||
| Interface | HANDLER, RS-232C, USBCDC, USBTMC, USBHOST | |||
| Model | ZX8815-5KV | ZX8815-3KV | ZX8815A-5KV | ZX8815A-3KV |
| Output Pulse Voltage | 100 V-5000 V 10 V step 5%±10 V | 100 V-3000 V 10 V step 5%±10 V | 100 V-5000 V 10 V step 5%±10 V | 100 V-3000 V 10 V step 5%±10 V |
| Inductance Test Range | >10 μH | >10 μH | >10 μH | >10 μH |
| Max Pulse Energy | MAX 0.25 J | MAX 0.09 J | MAX 0.25 J | MAX 0.09 J |
Display Waveform Display Area Display Content | 800×480 dot matrix graphic LCD 650×250 dot matrix Magnified display; setting parameters; standard and test waveforms; comparison results; file information, etc. | |||
| Waveform Sampling | Sampling Rate: 100MSa/s to 100kSa/s, 10 ranges Resolution: 8 bits Sampling Points: 6500 bytes | Sampling Rate: 200MSa/s to 200kSa/s, 10 ranges Resolution: 8 bits Sampling Points: 6500 bytes | ||
| Standard Waveform Sampling | Continuous Cycle, Single Cycle, Single Sample (Up to 32-sample averaging function) | |||
| Input Impedance | 10 MΩ | |||
| Test Speed | Up to 15 meas/sec | |||
| Pulse Count | Test pulse: programmable from 1 to 32 cycles Demagnetization pulses: programmable from 0 to 16 cycles | |||
| Display Waveform Measurement | Voltage, Time, Frequency | |||
| Trigger Mode | Internal / Manual (including foot control) / External / Bus | |||
| Comparison Method | AreaSize Comparison, DiffZone Comparison Corona Comparison, PhaseDiff Comparison | |||
| Area Repeatability | ±1% | |||
| Area Difference Repeatability | ±1% | |||
| Judgment Output | PASS/FAIL indication, LED display, Buzzer Alarm | |||
| Alarm Volume Selection | Long high beep, Long low beep, Single short beep, Double short beep, OFF | |||
| Memory | Internal: 200 sets External USB flash drive: 500 sets | |||
| Interface | HANDLER, RS-232C, USBCDC, USBTMC, USBHOST | |||

Unit: millimeters (mm)
Accessories & Options
ZX8815Y01R Test Cable
ZX8815Y01B Test Cable
ZX27Y02 Foot Switch
ZX28Y08B Small Inductor Interlayer Test Fixture
Software & Resources
Solutions
undefined
Customer Q&A
Q: How is the test voltage set?
A: There is no specific standard. Generally, 80% of the maximum voltage at which the waveform can still be clearly measured is taken as the test voltage. For example, if waveform distortion begins to appear in a good unit at 3000 V, the test voltage can be set to 2400 V. Adjustments may be made based on the actual defect rate distribution.
Q: How should the start and end points of a waveform be selected?
A: Typically, the peak of the first oscillation wave is taken as the start point, and the peak of the third or fourth oscillation wave is taken as the end point.
Q: How should the threshold range for area and area difference be configured?
A: There is no specific standard, it depends on the customer's internal criteria. It is recommended to test several good samples, record the maximum Area and Area Difference values, and slightly relax these values as the judgment thresholds. Then verify the detection capability using defective samples.
Q: What do the area and the area difference respectively represent?
A: ▪ Area: Within any specified A–B interval, the waveform area of the tested coil is integrated and compared with the area of the standard waveform over the same interval. The percentage difference between the two waveform areas relative to the standard waveform area in that interval serves as the judgment criterion, expressed as a percentage.
▪ Area Difference: Within any specified A–B interval, the Y-axis deviation between the test waveform and the standard waveform is calculated (the integration result represents the shaded area within the A–B interval). This difference area is compared with the standard waveform area over the same interval, with the judgment baseline set as a percentage.