ZX9502PV Series PV Hipot Tester

The ZX9502PV series PV hipot testers feature an 800×480 color LCD display and are powered by a mainstream high-performance 32-bit ARM processor. Specifically designed to meet the voltage testing requirements of the photovoltaic industry, they provide AC/DC test voltages of up to 10 kV. This series supports all current mainstream interface functions, enabling seamless data communication and remote control with a PC to build automated test systems.
  • File size: 327.0KB

  • File size: 2.0MB


Product Model

Model

Overview

Product Brochure

ZX9502PVPV Hipot Tester (AC/DC/IR); Max AC Current: 20 mA; Max AC Voltage: 10 kV.

Click to View

Functions & Advantages

● High-precision current measurement and fast voltage settling time

● Output voltage supports Normal Mode and Gradient Mode

● List sweep function for one-screen display of multi-step measurement results under different conditions

● Key lock function; automatically saves test conditions on power-off

● Open/Short Circuit Detection (OSC)

● Audible and visual alarm for defective products

● Fast discharge and arc detection functions

● Ground Fault Interrupter (GFI) for electric shock protection

● Independently settable voltage rise time, test time, and voltage fall time

● Each group stores up to 100 test steps; 60 groups internally,

240 groups on external USB flash drive

● Current baseline zeroing function

● Multiple interfaces: HANDLER, RS-232C, SCANNER, GPIB (optional)

● Instrument software upgrade via USB flash drive

 

Stable Output Unaffected by AC Power Fluctuations

Equipped with closed-loop feedback and linear amplifier technology for precise output voltage control — stable even under load variations!

Traditional hipot testers typically use conversion transformers for direct step-up, making the output voltage susceptible to AC line input fluctuations, which may lead to inaccurate test results. When distorted voltage is applied to the DUT, it may accelerate component degradation and create hidden quality issues. The ZX9502PV series employs closed-loop output voltage control, monitoring and adjusting the output in real time to ensure voltage fluctuation within ±(1% reading + 10 V) under various DUT load conditions. The internal independent regulated power supply design effectively reduces the impact of external grid disturbances, providing a solid foundation for testing.

 

High Precision and High-Speed Measurement

The ZX9502PV series delivers high-precision measurement capability with voltage measurement accuracy of ±(1.0% reading + 5 digits) and current measurement accuracy of ±(1.0% reading + 5 digits). Equipped with auto-ranging, it achieves high-speed judgment with a minimum test time of 0.3 seconds, ensuring fast and accurate measurements.

◆  Current Zeroing Function

During testing, changes in ambient temperature and humidity as well as different cable routing can cause stray currents from distributed capacitance in the test leads and fixtures, resulting in a baseline reading under open-circuit conditions. The current zeroing function eliminates such stray current interference with one-touch, further improving measurement accuracy and ensuring reliable insulation performance evaluation.

◆  Arc Detection Function

Specifically designed to capture transient discharge pulses in the high-voltage test loop, filtering out normal current and responding only to high-speed current variations. Due to low-pass filter characteristics and the inherent randomness of arc magnitude, this function can only provide an approximate assessment of localized arcing severity.

 

Precise Time Control of Applied Voltage

Features wide-range voltage rise/fall time control! Prevents adverse effects of over-spec voltage surges on the DUT.

The ZX9502PV series supports both rise time and fall time control in AC withstand, DC withstand, and insulation resistance tests. Unlike traditional instruments that only allow short fixed rise/fall times, the ZX9502PV allows rise or fall times to be set up to 999.9 seconds. This capability is especially useful for large capacitive loads or DUTs sensitive to voltage transients, effectively preventing insulation damage or component aging caused by transient overvoltage and ensuring tests are both rigorous and safe.

◆  Rise Time Control Function

Users can set the time for voltage to rise from 0 to the set value based on DUT characteristics, adjustable from 0.1 s to 999.9 s (can be disabled).

◆  Fall Time Control Function

After the test, users can set the time for voltage to smoothly decrease from the set value to 0, adjustable from 0.1 s to 999.9 s (can be disabled).

 

Active Safety Protection Mechanisms

◆  Ground Fault Interrupter (GFI)

Monitors ground return current in real time. If the threshold of 0.5 mA is exceeded, the instrument forcibly cuts off high-voltage output and exits the test state. Even if the operator disables this function, the instrument will still automatically trigger protection when the ground current exceeds 70 mA, ensuring no serious injury to the operator. Provides active, reliable safety protection for operators.

◆  Interlock Switch (INTERLOCK)

INTERLOCK is used for multi-unit interconnection, preventing damage caused by simultaneous output from multiple instruments due to misoperation. When the interlock switch is cut off, the instrument forcibly inhibits high-voltage startup. In other words, the interlock loop being closed is a prerequisite for the instrument to output high voltage.

◆  Forced Fast Discharge Function

After DC withstand or insulation resistance testing, the instrument automatically discharges high-voltage charge stored in capacitive loads to a safe voltage level, preventing residual electric shock hazards.

◆  Open/Short Circuit Detection (OSC)

Pre-tests the test loop before high-voltage application, capable of detecting distributed impedance currents above 100 pF to confirm reliable connections, preventing misjudgment caused by open test leads or poor pin contact, and ensuring accurate and safe high-voltage testing.

 

Multi-Step Results on One Screen

◆  List Sweep Function

Users can freely edit multiple measurement steps based on test requirements, with each step independently configurable for test voltage, test time, upper/lower limits, and other parameters. Enables one-screen display of measurement results from different steps and conditions.

 

Wide Range of Applications

◆  Power Equipment and Grid Maintenance

◆  Cable and Transformer Manufacturing

◆  New Energy Sector

◆  Industrial Electrical Equipment

◆  Rail Transit and Aviation

◆  Laboratories and Certification Bodies

Parameters & Dimensions

ModelZX9502PV
Withstanding Voltage Test
Output VoltageAC0.5 kV–10 kV ±(1.0% of reading + 5 digits) (50 Hz and 60 Hz selectable)
DC0.5 kV–10 kV ±(1.0% of reading + 5 digits)
Voltage Regulation Rate≤(1.0% + 10 V)(Rated Power)
Current Test RangeAC0.1 mA–20 mA (Max 200 VA AC)
DC0.01 mA–10 mA (Max 60 VA DC)
Test Accuracy±(1.0% of reading + 5 digits)
Discharge FunctionAutomatic discharge after test (DCW)
Insulation Resistance Test
Output Voltage0.5 kV–2.5 kV ±(1.5% of reading + 5 V)
Resistance Measurement Range5 MΩ-50 GΩ.
Discharge FunctionAutomatic discharge after test
Arc Detection
Measurement RangeACLevels 1–9 (Level 9 is the most sensitive)
DCLevels 1–9 (Level 9 is the most sensitive)
General Parameters
Memory120 sets, 100 steps per set
Voltage Rise Time0.1 s to 999 s
Voltage Fall Time0.1 s to 999 s
Voltage Dwell Time0.1 s to 99.9 s (DC only)
Test Time Setting0.3 s to 999 s
Interface
StandardRS-232C,HANDLER, REMOTE I/O, USBHOST
OptionalGPIB, SCAN

Unit: millimeters (mm)

Accessories & Options

ZX95Y02B Test Cable

ZX95Y02B Test Cable

ZX95Y02R Test Cable

ZX95Y02R Test Cable

ZX95Y01R High-Voltage Test Probe

ZX95Y01R High-Voltage Test Probe

ZX95Y03B High-Voltage Test Rod

ZX95Y03B High-Voltage Test Rod

ZX95Y03R High-Voltage Test Rod

ZX95Y03R High-Voltage Test Rod

Software & Resources

Solutions

undefined

undefined

Customer Q&A

Q: What should be done if arcing or sparking occurs during measurement?

A: This can be mitigated by adding ferrite cores to the test leads and ensuring reliable grounding of the instrument chassis.

Q: Why is there a current reading or an abnormally low resistance value during an open-circuit test after the fixture is connected?

A: Ensure that the test fixture itself has good insulation. If the reading returns to normal after removing all test leads and performing an open-circuit test, the issue is most likely due to insufficient insulation of the test fixture. For minor current offset issues, the baseline current can be cleared using the zeroing function.

Total 2 < 1 >

Enter information to retrieve data.

By systematically retrieving and analyzing relevant input information, comprehensive and specialized data are obtained to support or meet specific needs or objectives.

Confirm
%{tishi_zhanwei}%